Example
>ctext
info : / FIN ?
>
6.5.2 CDATE command
Use the
CDATE
command to enter a date to calibration information field. Set the calibration
date in format YYYYMMDD.
Example
>cdate 20080320
>
6.6 Analog output adjustment
When adjusting the analog output the analog output is forced to the following values:
Voltage output: 10 % and 90 % of the range
To measure voltage, connect HMP155 to a calibrated voltage meter.
To change HMP155 to the adjustment mode, hold down the
ADJ
button until a
green LED lights up.
You cannot use the serial line command
ACAL
with HMP155 passive output
version.
Type the
ACAL
command and type the multimeter reading for each case. Continue by pressing
ENTER
. When a channel is specified, only the specified analog output channel is adjusted.
6.6.1 ACAL command
Use the
ACAL
command for analog calibration and adjustment.
Chapter 6 – Calibration and adjustment
77