6.4.5 CRH command
Use the
CRH
command for one-point or two-point serial line relative humidity calibration and
adjustment.
CRH
changes the user adjustment parameters of relative humidity. See also
The
CRH
command can only be used in the adjustment mode.
Example of two-point adjustment:
>crh
RH : 16.6675 1. ref ?
RH : 16.4978 1. ref ?
RH : 16.3956 1. ref ? 11.25
Press any key when ready ...
RH : 11.25 Ref2 ?
RH : 75.45 Ref2 ?
RH : 75.57 Ref2 ?
RH : 75.55 Ref2 ?
RH : 75.59 Ref2 ? 75.5
OK
OK
indicates that the adjustment has succeeded and the new calibration coefficients are
calculated and stored.
6.4.6 FCRH command
Use the
FCRH
command for relative humidity calibration after a sensor change.
>fcrh
RH : 25.19 Ref1 ? 11.3
Press any key when ready ...
RH : 70.02 Ref2 ? 75.5
OK
>
OK indicates that the adjustment has been successful.
6.5 Typing calibration information
In the adjustment mode, you can only use the
CTEXT
and
CDATE
commands. To change to the
adjustment mode, hold down the
ADJ
button until the green LED lights up.
6.5.1 CTEXT command
Use the
CTEXT
command to enter text to the calibration information field.
HMP155 User Guide
M210912EN-D
76
Содержание HUMICAP HMP155
Страница 1: ...M210912EN D User Guide Vaisala HUMICAP Humidity and Temperature Probe HMP155 ...
Страница 8: ...HMP155 User Guide M210912EN D 6 ...
Страница 90: ...7 7 3 Radiation shield dimensions Figure 17 DTR503 and DTR13 dimensions HMP155 User Guide M210912EN D 88 ...
Страница 94: ......
Страница 98: ...HMP155 User Guide M210912EN D 96 ...
Страница 100: ......
Страница 101: ......
Страница 102: ...www vaisala com ...