
4.4 Measurement commands
4.4.1 R command
Use the
R
command to start the continuous output of measurements. The command starts a
temporary RUN mode. To change to the permanent RUN mode, use the
SMODE
command.
>r
RH= 33.0 %RH T= 22.1 'C
>
If a value is too long to fit in the output or if there is an error in the quantity output, the value is
displayed with stars (*).
To adjust the output format, use the
FORM
command.
To change the output interval, use the
INTV
command.
4.4.2 S command
Use the
S
command to end the RUN mode. After the
S
command, you can use any command.
To stop the output, you can press the
Esc
or reset the probe.
To change the default power-up operation mode, see
4.4.3 INTV command
Use the
INTV
command to set the output interval for the RUN mode. The default value is 2
seconds.
INTV xxx yyy
Parameter
Description
xxx
Output interval (0 ... 255) where 0 produces an output whenever a new
measurement is made. Interval 1 s produces repeats of the previous
measurement if no newer measurement is available.
yyy
Unit (s, min, or h)
HMP155 User Guide
M210912EN-D
42
Содержание HUMICAP HMP155
Страница 1: ...M210912EN D User Guide Vaisala HUMICAP Humidity and Temperature Probe HMP155 ...
Страница 8: ...HMP155 User Guide M210912EN D 6 ...
Страница 90: ...7 7 3 Radiation shield dimensions Figure 17 DTR503 and DTR13 dimensions HMP155 User Guide M210912EN D 88 ...
Страница 94: ......
Страница 98: ...HMP155 User Guide M210912EN D 96 ...
Страница 100: ......
Страница 101: ......
Страница 102: ...www vaisala com ...