4.8.1 # command
Use the
#
command to force the settings 19200 baud, no parity, 8 data bits, 1 stop bit during
the first 3 seconds after power-up.
The settings last for one session. Next time you connect, HMP155 reverts to the settings set
with the
SERI
command.
More information
‣
4.8.2 AMODE command
Use the serial line to select and scale the analog output quantities. Connect HMP155 to the PC
and open the terminal connection between your computer and the sensor.
HMP155 must be in the adjustment mode for the
AMODE
command to work. To
switch the probe to the adjustment mode, hold down the
ADJ
button until a
green LED lights up.
Check the analog output modes with the
AMODE
command.
Table 26 Analog output modes
Analog output mode
Description
3
0 … 1 V
4
0 …5 V
5
0 … 10 V
Analog output calibration is valid only for the factory settings. If you change the
analog output mode with
AMODE
command, you must continue by entering the
ACAL
command, see
.
Example
>amode 3 3
Ch0 : 0 ... 1V
Ch1 : 0 ... 1V
>
Chapter 4 – Operation
53