2. Carefully remove the filter from the probe. Holding the filter by the plastic part, rotate the
filter counter-clockwise and pull it out.
Do not touch the sensor heads.
CAUTION!
1
Filter
2
O-ring
3
HUMICAP humidity sensor
4
Pt100 temperature sensor
3. After removing the filter, check the O
‑
ring and change it if necessary.
4. Install the new filter carefully without delay.
5. Install the probe back inside the radiation shield.
5.1.3 Changing HUMICAP sensor
You can change INTERCAP, HUMICAP180R, and HUMICAPR2 sensor yourself.
Never replace INTERCAP with HUMICAP180R, or INTERCAP with
HUMICAPR2, or the other way round.
You can replace HUMICAP180R with HUMICAPR2, and the other way round.
CAUTION!
Chapter 5 – Maintenance
63