Figure 12 HMP155 adjustment buttons
1
Protective cover (unscrewed)
2
Protective plug (lifted up)
3
Down button
4
ADJ button
5
Up button
There is a 2-color indicator LED (green and red) under the protective plug and next to the
Down
button.
If you use a warmed probe (HMP155 active output version option), probe heating
is interrupted when you press the
ADJ
button. Allow sufficient time for the probe
to reach ambient temperature before starting the adjustment procedure.
Fixed pressure compensation value of 1.01325 bar is used when in adjustment
mode.
6.2 Push-button calibration
You can calibrate and adjust HMP155 with the adjustment buttons on the probe. Carry out the
two-point humidity adjustment by using 2 relative humidity references: for example, saturated
salt points 11 % RH (LiCl) and 75 % RH (NaCl). The difference between the humidity reference
points must be at least 30 %RH. The difference between the temperature reference points
must be at least 30 °C.
Chapter 6 – Calibration and adjustment
67