![Vaisala CAB100 Скачать руководство пользователя страница 118](http://html2.mh-extra.com/html/vaisala/cab100/cab100_installation-manual_3171839118.webp)
Input type
Current loop
Analog voltage
Overload protection
40 mA max. (reverse-polarity
protected)
±24 VDC max. (reverse-polarity
protected)
9.2.3 Vaisala vNet PoE Data Logger Interface specifications
Table 18 vNet operating environment
Property
Description/Value
Operating temperature
−25 … +70 °C (−13 … +158 °F)
Storage temperature
−40 … +85 °C (−40 … +185 °F)
Operating humidity
0 … 90 %RH (non-condensing and not to exceed a
mixing ratio of 38.5 g/kg)
Table 19 vNet inputs and outputs
Property
Description/Value
Ethernet connectivity
IEEE 802.3af (Class 1)
, bandwidth 10Base-T
Connectivity cable
Category 5/5e
RJ-45 connector
1.83 m (6 ft)
Operating voltage
12 … 30 VDC
Plugs into vNet jack labeled 12 V
Output voltage
Model CDL-VNET-LPC
Nominal: 15 VDC
Maximum: 350 mW
Power consumption
Model CDL-VNET-LPC
Typical: 900 mW
Maximum: 1.35 W
Power supply
North America
12 VDC / 0.5 A max. out
120 VAC in
International
12 VDC / 1.66 A max. out
100 … 240 VAC in
1)
Max. PSE power reservation is 4.00 W.
2) Optional for use without PoE.
3) Included but not required when using PoE.