The command does not affect the following configurations:
• Device identification information
• Factory calibration information
The command does not affect connected devices, such as laser particle counter (LPC) and
the temperature and humidity (TH) probe.
Example response:
Loading default configuration... OK!
Write the default configuration to non-volatile memory by using "write"
command.
$
Use the
write
command to save the configuration.
write
--really
5.5 Printing LPC information (
lpc
command)
To display information about the laser particle counter (LPC), use the
lpc
command.
lpc
--info
Table 20 Parameters for
lpc
command
Parameter
Value
Mandatory
Description
--info
Fixed
Yes
Shows LPC module information:
• Serial number
• Firmware version
The LPC must be connected to the system.
Querying LPC information:
Chapter 5 – CLI commands and parameters
31
Содержание AQT530
Страница 1: ...M212572EN C Configuration Guide Vaisala Air Quality Transmitter AQT530...
Страница 62: ......
Страница 63: ......
Страница 64: ...www vaisala com...