Address
RW
Type
Unit
Register
count
Description
0x0002
R
int16
ppb
1
Carbon monoxide (CO) concentration in parts
per billion (ppb) with linear correction
• For a value without linear correction, use
register 0x006E
• For µg/m
3
value, use register 0x0068
0x0004
R
int16
ppb
1
Hydrogen sulfide (H
2
S) concentration in parts
per billion (ppb) with linear correction
• For a value without linear correction, use
register 0x0070
• For µg/m
3
value, use register 0x0069
0x0005
R
int16
ppb
1
Trioxygen (ozone) (O
3
) concentration in parts
per billion (ppb) with linear correction
• For a value without linear correction, use
register 0x0071
• For µg/m
3
value, use register 0x006A
0x0006
R
int16
ppb
1
Nitric oxide (NO) concentration in parts per
billion (ppb) with linear correction
• For a value without linear correction, use
register 0x0072
• For µg/m
3
value, use register 0x006B
0x0008
R
int16
0.1 μg/m
3
1
Particulate matter PM
2.5
with linear correction
• For a value without linear correction, use
register 0x0073
0x0009
R
int16
0.1 μg/m
3
1
Particulate matter PM
10
• For a value without linear correction, use
register 0x0074
0x000A
R
int16
0.1 °C /
0.1 °F
1
Air temperature (HMP110) in Celsius or
Fahrenheit, depending on configuration
Current configuration available from register
0x001C
0x000B
R
int16
0.1 %RH
1
Air humidity (HMP110)
0x000C
R
int16
0.1 hPa
1
Air pressure from an on-board analog pressure
sensor
0x0013
R
uint16
-
1
Calibration year
0x0014
R
uint16
-
1
Calibration month
0x0015
R
uint16
-
1
Calibration day
AQT530 Configuration Guide
M212572EN-C
46