Parameter
Type
Range
Default
Unit
Description
no2_span
uint8
1 ... 255
100
0.01
NO
2
sensor span
correction (1 ... 255 =
0.01 ... 2.55)
so2_zero
int16
−10000 ... 10000
0
ppb
SO
2
sensor zero
correction
so2_span
uint8
1 ... 255
100
0.01
SO
2
sensor span
correction (1 ... 255 =
0.01 ... 2.55)
h2s_zero
int16
−10000 ... 10000
0
ppb
H
2
S sensor zero
correction
h2s_span
uint8
1 ... 255
100
0.01
H
2
S sensor span
correction (1 ... 255 =
0.01 ... 2.55)
o3_zero
int16
−10000 ... 10000
0
ppb
O
3
sensor zero
correction
o3_span
uint8
1 ... 255
100
0.01
O
3
sensor span
correction (1 ... 255 =
0.01 ... 2.55)
Table 16 Linear correction parameters - particles
Parameter
Type
Range
Default
Unit
Description
pm1_zero
int16
−10000 ... 10000
0
0.1 μg/m
3
PM
1
offset (zero)
correction
(−10000 ... 10000 =
−1000.0 ... 1000.0)
pm1_span
int16
1 ... 10000
1000
0.001
PM
1
span (gain)
correction (1 ... 10000
= 0.001 ... 10.000)
pm25_zero
int16
−10000 ... 10000
0
0.1 μg/m
3
PM
2.5
offset (zero)
correction
(−10000 ... 10000 =
−1000.0 ... 1000.0)
pm25_span
int16
1 ... 10000
1000
0.001
PM
2.5
span (gain)
correction (1 ... 10000
= 0.001 ... 10.000)
pm10_zero
int16
−10000 ... 10000
0
0.1 μg/m
3
PM
10
offset (zero)
correction
(−10000 ... 10000 =
−1000.0 ... 1000.0)
pm10_span
int16
1 ... 10000
1000
0.001
PM
10
span (gain)
correction (1 ... 10000
= 0.001 ... 10.000)
AQT530 Configuration Guide
M212572EN-C
26