Parameter
Type
Range
Default
Unit
Description
rs485_databits
uint8
7, 8
8
bits
RS-485 data bits
rs485_parity
Text
•
N
= None
•
E
= Even
•
O
= Odd
E
−
RS-485 parity
Case insensitive
rs485_stopbits
uint8
1, 2
1
bits
RS-485 stop bits
Table 25 Temperature unit parameter
Parameter
Type
Range
Default
Unit
Description
tempunit
uint8
•
0
= °C
•
1
= °F
0
−
Temperature unit for
Modbus, ASCII CSV,
and CLI interfaces
Table 26 Linear correction parameters - gases
Parameter
Type
Range
Default
Unit
Description
co_zero
int16
−10000 ... 10000
0
ppb
CO sensor zero
correction
co_span
uint8
1 ... 255
100
0.01
CO sensor span
correction (1 ... 255 =
0.01 ... 2.55)
no_zero
int16
−10000 ... 10000
0
ppb
NO sensor zero
correction
no_span
uint8
1 ... 255
100
0.01
NO sensor span
correction (1 ... 255 =
0.01 ... 2.55)
no2_zero
int16
−10000 ... 10000
0
ppb
NO
2
sensor zero
correction
no2_span
uint8
1 ... 255
100
0.01
NO
2
sensor span
correction (1 ... 255 =
0.01 ... 2.55)
so2_zero
int16
−10000 ... 10000
0
ppb
SO
2
sensor zero
correction
so2_span
uint8
1 ... 255
100
0.01
SO
2
sensor span
correction (1 ... 255 =
0.01 ... 2.55)
h2s_zero
int16
−10000 ... 10000
0
ppb
H
2
S sensor zero
correction
h2s_span
uint8
1 ... 255
100
0.01
H
2
S sensor span
correction (1 ... 255 =
0.01 ... 2.55)
o3_zero
int16
−10000 ... 10000
0
ppb
O
3
sensor zero
correction
Chapter 5 – CLI commands and parameters