Action
CLI command and
parameter
CLI reference
Modbus register
addresses
Get ug/m
3
converted
values
meas
--metric
and changing
measurement output unit
Modbus register
addresses are available
for both metric (ug/m
3
)
and parts per billion
(ppb) values
Reset device
reboot
--really
Restarting device
( reboot command)
(page 35)
0x00FA
To use the CLI commands, establish a local maintenance connection to AQT530 over RS-232.
More information
‣
Overview of CLI commands and parameters (page 28)
‣
Modbus register addresses for AQT530 (page 45)
3.3 Measurement output
The AQT530 measurement data is available over RS-485 by means of Modbus (ASCII and
RTU modes), or alternatively in ASCII CSV format.
The measurement data can be monitored also through CLI.
The following tables present the measurements, their units, and resolution.
Set of available gases depends on gas cell setup.
Particle measurement results are included when an LPC device is installed and enabled.
Table 5 Temperature
Value
Modbus
CLI
ASCII CSV
Air temperature
0x000A
: 0.1 °C / 0.1 °F
•
meas
: °C / °F
•
meas
--metric
: °C / °F
•
meas
--ppb
: °C / °F
°C / °F
Table 6 Humidity
Value
Modbus
CLI
ASCII CSV
Air humidity
0x000B
: 0.1 %RH
•
meas
: %RH
•
meas
--metric
: %RH
•
meas
--ppb
: %RH
%RH
AQT530 Configuration Guide
M212572EN-C
14