Field
Description
Example
Particles
Particle measurement results in µg/m
3
,
separated by a comma:
• PM
1
, PM
2.5
, PM
10
The particle measurement results are included
when an LPC device is installed and enabled
0.1,1.1,1.9
Config
Device setup represented with symbols
separated by colon (:):
• T = Air temperature (included always)
• H = Air humidity (included always)
• P = Air pressure (included always)
• NO
2
, SO
2
, CO, H
2
S, O
3
, NO (included
according to gas cell setup)
• PM
1
, PM
2.5
, PM
10
(included when LPC is
installed and enabled)
Order matches order of results in fields
Conditions, Gases, and Particles
T:H:P:NO2:CO:O3:NO:PM1:PM2.5:PM1
0
Uptime
Time from last reboot in seconds
2735641
1)
The temperature unit can be configured with CLI using
tempunit
parameter. For example, setting Celsius units with
set
tempunit
=
0
.
Example of CSV ASCII data message with 4 gas sensors and LPC
2022-01-22T07:37:38,22.3,24.1,999.3,0.182,2.920,0.575,0.140,0.1,1.1,1.9,T:H:P:N
O2:CO:O3:NO:PM1:PM2.5:PM10,3185
2022-01-22T07:38:38,22.3,24.1,999.3,0.170,2.921,0.551,0.131,0.1,1.1,1.9,T:H:P:N
O2:CO:O3:NO:PM1:PM2.5:PM10,3245
2022-01-22T07:39:38,22.3,24.1,999.3,0.159,2.919,0.527,0.123,0.1,1.1,1.9,T:H:P:N
O2:CO:O3:NO:PM1:PM2.5:PM10,3305
Example of CSV ASCII data message with 4 gas sensors, without LPC
2022-01-22T08:07:38,22.3,24.1,999.4,0.108,2.926,0.416,0.084,T:H:P:NO2:CO:O3:NO,
4983
2022-01-22T08:08:38,22.3,24.0,999.4,0.101,2.927,0.402,0.079,T:H:P:NO2:CO:O3:NO,
5043
2022-01-22T08:09:38,22.3,24.0,999.4,0.095,2.927,0.389,0.074,T:H:P:NO2:CO:O3:NO,
5103
Example of CSV ASCII data message with LPC only
2022-01-22T07:40:38,22.4,24.1,999.3,0.1,1.1,1.9,T:H:P:PM1:PM2.5:PM10,3364
2022-01-22T07:41:38,22.4,24.1,999.3,0.1,1.1,1.9,T:H:P:PM1:PM2.5:PM10,3424
2022-01-22T07:42:38,22.4,24.1,999.3,0.1,1.1,1.9,T:H:P:PM1:PM2.5:PM10,3484
Chapter 4 – CLI operation