page 34 of 58
Shutdown
Has the pump been exposed to condensate?
-
Allow the pump to continue to run for a few minutes
with the rotary vane pump’s gas ballast valve open be-
fore switching off.
Long-term:
- Separate the pump from the application.
- Drain condensates (catchpots of the separator at the
inlet and of the oil mist filter).
- Flush the pump with dry nitrogen.
- Carry out oil change (see “Oil change”, pg. 35).
- Fill the pump completely with new oil (to above the
”max.” mark!).
Attention: Before restarting the pump, drain oil to the
maximum oil level (”max.” mark)!
- Close the manual gas ballast valve.
- Close inlet and outlet ports (e.g., with transport caps or
blind flanges).
- Store the pump under dry conditions.
- Carry out oil change and maintenance prior to use if
the pump has been stored for longer than one year.
Note:
When transporting a pump containing oil, take great
care not to tilt the pump to an angle that would allow oil to
make its way into the filter element. Drain oil if necessary.
NOTICE