page 7 of 58
General information
Remove all packing material from the packing box. Re-
move the product from its packing-box and retain all pack-
aging until the equipment is inspected and tested. Re-
move the protective caps from the inlet and outlet ports
and retain for future use. Inspect the equipment promptly
and carefully.
If the equipment is damaged, notify the supplier and the
carrier in writing within three days. Retain all packing ma-
terial for inspection. State the item number of the product
together with the order number and the supplier’s invoice
number. Failure to check and give notice of damage will
void any and all warranty claims for those deficiencies.
Replace the protective caps, if the equipment is not used
immediately. Store the equipment in dry and non-corrosive
conditions (see also “Technical data”, pg. 19).
+
Do not use any damaged equipment.
• Check the pump‘s oil level.
• The pump is delivered without oil filling.
Fill with oil
before operating the pump
• Aging of the oil (as indicated by darker color compared
to new oil, strange odor of the oil, particles in the oil, or
contamination) necessitates an oil change.
• Use the handle when moving the pump.
Intended use
+
Do not use the pump or any system parts on humans
or animals.
+
Ensure that the individual components are only con-
nected, combined and operated according to their de-
sign and as indicated in the instructions for use.
Use only
original manufacturer’s spare parts and
NOTICE