07
Product Installation
Installation environment
--------------------------------------------
The freezer is only for indoor use, with second pollution level and second
type of over-voltage
Ambient temperature is within 16°C-32°C, and recommended temperature is
within 18°C-25°C. If necessary, air conditioning system should be used.
Environmental humidity: lower than 80%Rh.
Avoid heavy dust.
Avoid mechanical sway or vibration.
The altitude of working position of freezer is lower than 2000 m.
Input voltage: 220 V±10%.
The freezer is sensitive to ambient temperature. When installed in other
environments than the one aforesaid, the freezer can't operate normally. Please
improve the environment before use.
It is prohibited to install the freezer in an open environment. When the freezer is
wet by rain, it may cause electric leakage or electric shock.
Installation Site
---------------------------------------------------------
Do not install the device in a narrow and confined space. Install it in a place that
can ensure its normal access to avoid damage to storage goods by maintenance
difficulties and delayed repair in case of device failure.
The installation floor must be solid, flat, non-combustible and able to bear the
weight of the device during operation.
Provide good ventilation and avoid direct sunlight.
Each device needs to use a separate power outlet. Please ensure that the power
outlet should withstand a current of ≥10 A, and the plug and power outlet are
firmly connected.
Check the working voltage before use. In areas with unstable voltage, a voltage
stabilizer suitable for device load should be considered for voltage stabilization,
and the power of the voltage stabilizer should be greater than 4000 W to ensure
that the input voltage requirements in the installation environment can be met.