14
Manual export of USB data:
when the key is unlocked and no file is being
generated, connect the USB flash disk and press the key for 3s until the digital
tube displays "d01". Press the or
key to adjust "d00
~
d12". Press the
key,
and the file generated this time (d00) or the PDF file with the data recorded in the
forward month (1-12) is cancelled inside the USB flash disk.
Note:
When the digital tube flashes and displays "LoF", the recorder is not
started. Press down the and
keys simultaneously for 3 seconds, "LoF"
disappears and the recorder starts.
Independent closing of the upper or lower chambers:
it is used for inter-chamber
deactivation, or separate frosting.
In the normal operation mode, first unlock the panel, and press and hold the keys
and together to close the upper chamber. The upper digital tube flashes and
displays "off".
In the normal operation mode, first unlock the panel, and press and hold the keys
and
together to close the lower chamber. The lower digital tube flashes and
displays "off".
To modify the time of the data recorder system:
The system time can be set when the machine is powered on again after power
failure. The digital tube displays "n" after the system completes the self test. If there is
no operation for 60 seconds, it will automatically exit and enter the normal display.
Menu
Menu Description
Setting Range
Note
n
Set the time of recorder
module - year
10~50
22 represents 2022
y
Set the time of recorder
module - month
1~12
--
r
Set the time of recorder
module - date
01~31
--
S
Set the time of recorder
module - hour
00~23
--
F
Set the time of recorder
module - minute
00~59
--
Pt
Printing interval
0~240
20min
SCY
Temperature data recording
cycle
0~240
0: shielded recorder
10min