15
Press the
key to switch parameters n-y-r-s-F-Pt-SCY
Press the key or the
key to set the parameter value. After setting, press and
hold
to save the parameter.
Note: Operation to modify the time and recording interval will cause the recorder
data to be cleared, and the data recorder will be restarted. The digital tube will
display "LoF". To eliminate it, refer to Recorder Startup Operations. When the digital
tube flashes and displays "LoF", the recorder is not started. Press down the
and
keys simultaneously for 3 seconds, "LoF" disappears and the recorder starts
View the maximum and minimum values
The MAX1 and MAX2 parameters in the user menu, as well as the MIN1 and MIN2
parameters, display the maximum and minimum temperature values within the
current time period
Battery voltage check
Press down the keys
+
, the digital tube displays the battery voltage
Function Settings
User Parameter Setting
Unlocking:
In the normal operation state, press and hold the and
keys for
3 seconds, the upper digital tube displays "PS1"and the lower digital tube displays
the parameter code "0000". Press the key, enter the password "0005" (When
entering the user menu password, enter "0099" to restore the key lock password
to the default "0005".),and it will be unlocked. After unlocking, press and hold the
key for 3 seconds, and the digital tube displays the parameter code "MAX", thus
entering the setting to adjust the parameters.