09
First Power On
-----------------------------------------------------------
When using the device for the first time, please follow these steps:
1. After the device is placed, leveled and cleaned, it needs to be left for more
than 24 hours before it is connected to the power supply to ensure the normal
operation of the device.
2. Under no-load condition, connect the power cord to a special socket with
appropriate specifications.
3. After turning on the power, observe whether the machine is running normally.
4. Check whether the operating temperature of the device reaches the required
temperature, observe whether the device is normally on and off for more
than 24 hours, and put a small amount of articles in the storage room after
confirming that the performance is normal.
5. Please store items in batches, which is not exceeding 1/3 of the box volume at
a time. The next batch of goods can only be put in after the device is shut down
for more than 12 hours.
6. Try not to open the door during the cooling process, otherwise it will cause the
temperature to rise.
Operation After Power Failure
--------------------------------------
The device has the memory function for the set value. When the power is restored
after the power failure, the device will continue the operation before the power
failure.
Warning
Once the power of the freezer is cut off, it will take more
than five minutes to turn it on again to avoid damage to the
compressor or system.
When the equipment is not used for a long time, the power
should be cut off and the battery should be turned off at the
same time (for projects with batteries).
Please ensure that children will not approach the medical
freezer and use it as a toy if it is left unattended in an
unsupervised area for a long period.