16
1. Scroll through the parameters using the
key, and the display order is
MAX1 -MIN1-MAX2-MIN2-CLR-SEt1- H01-L01-SEt2-H02-L02-n-y-r-S-F-Pt-tH1-P1-
PS1-b1-b2
2. Increase or decrease the value using the and
keys;
3. Press the
key to temporarily store the modified value and return to the display
parameters
4. If you need to modify other parameters, repeat steps 1~3.
5. Press and hold the
key for more than 3 seconds to store the modified
parameter and return to the displayed parameter category
Serial
Number Menu
Parameter
Range
Recommended
Setting Value
Note
1
MAX1
---
Upper maximum temperature in the
current time period
2
MIN1
---
Upper minimum temperature in the
current time period
3
MAX2
---
Lower maximum temperature in the
current time period
4
MIN2
---
Lower minimum temperature in the
current time period
5
CLR
0~1
1
0: Clear the record of the maximum and
minimum temperature in the current
time period
1: The record of the maximum and
minimum temperature in the current
time period will not be cleared
6
SEt1
MD-25L series: -10~-25°C
VS
-30L series: -10~-30°C
VS
-40L series: -20~-40°C
-25 °C
-30 °C
-40 °C
Upper chamber temperature setting
value
7
H01
0.0~10.0
5
Upper deviation on high temperature
alarm of the upper chamber
8
L01
0.0~10.0
5
Lower deviation on low temperature
alarm of the upper chamber
9
SEt2
MD-25L series: -10~-25°C
VS
-30L series: -10~-30°C
VS
-40L series: -20~-40°C
-25 °C
-30 °C
-40 °C
Lower chamber temperature setting
value
10
H02
0.0~10.0
5
Upper deviation on high temperature
alarm of the lower chamber
11
L02
0.0~10.0
5
Lower deviation on low temperature
alarm of the lower chamber