1. Scroll through the parameters using the 

 key, and the display order is

MAX1 -MIN1-MAX2-MIN2-CLR-SEt1- H01-L01-SEt2-H02-L02-n-y-r-S-F-Pt-tH1-P1-
PS1-b1-b2

2. Increase or decrease the value using the   and 

 keys;

3. Press the 

 key to temporarily store the modified value and return to the display

parameters

4. If you need to modify other parameters, repeat steps 1~3.
5. Press and hold the 

 key for more than 3 seconds to store the modified

parameter and return to the displayed parameter category

Serial 

Number Menu

Parameter 

Range

Recommended 

Setting Value

Note

1

MAX1

---

Upper maximum temperature in the 

current time period

2

MIN1

---

Upper minimum temperature in the 

current time period

3

MAX2

---

Lower maximum temperature in the 

current time period

4

MIN2

---

Lower minimum temperature in the 

current time period

5

CLR

0~1

1

0: Clear the record of the maximum and 

minimum temperature in the current 

time period

1: The record of the maximum and 

minimum temperature in the current 

time period will not be cleared

6

SEt1

MD-25L series: -10~-25°C

VS

-30L series: -10~-30°C

VS

-40L series: -20~-40°C

-25 °C

-30 °C

-40 °C

Upper chamber temperature setting 

value

7

H01

0.0~10.0

5

Upper deviation on high temperature 

alarm of the upper chamber

8

L01

0.0~10.0

5

Lower deviation on low temperature 

alarm of the upper chamber

9

SEt2

MD-25L series: -10~-25°C

VS

-30L series: -10~-30°C

VS

-40L series: -20~-40°C

-25 °C

-30 °C

-40 °C

Lower chamber temperature setting 

value

10

H02

0.0~10.0

5

Upper deviation on high temperature 

alarm of the lower chamber

11

L02

0.0~10.0

5

Lower deviation on low temperature 

alarm of the lower chamber

Содержание VS -25L Series

Страница 1: ...EN 61326 1 2013 Product implementation standards EN 61010 1 2010 A1 2019 EN 61010 2 011 2017 USER MANUAL MODEL VS 25L518 VS 30L518 VS 40L518 Biomedical...

Страница 2: ...r Use 06 4 Product Installation 07 5 Product Composition and Overview 11 6 Operating Instructions 12 7 Function Settings 15 8 Troubleshooting and Maintenance Services 20 9 Specifications 22 10 Packing...

Страница 3: ...rature data can be traced and the test data and the temperature curve inside the box can be output at any time Efficient Refrigeration Energy Saving Mute Vacc Safe is a world renowned compressor brand...

Страница 4: ...icians Be sure to place the freezer firmly on a solid and flat ground If the ground is not stable or the place is not suitable the freezer might be overturned or people will be injured If the power co...

Страница 5: ...en storing toxic harmful or radioactive materials please use a freezer in a safe area Improper use may cause harm to human health or the environment When the freezer is not used for a long time unplug...

Страница 6: ...ining water on the freezer If the object falls it may cause personal injuries and the outflow of water will reduce the insulation and cause leakage or electric shock Do not insert metal objects such a...

Страница 7: ...or personal injury Please do not use the door handle to lift or carry the equipment in case of injury to the freezer or personal injury The refrigeration circuit must not be damaged Do not use electri...

Страница 8: ...during the test After the test line is led out it is necessary to plug the test hole with insulating material again otherwise the temperature in the box may not reach the set temperature value and co...

Страница 9: ...it may cause electric leakage or electric shock Installation Site Do not install the device in a narrow and confined space Install it in a place that can ensure its normal access to avoid damage to st...

Страница 10: ...vice may not be able to operate normally Please improve the environment before using the device the device is working in intermittent operation Ambient storage and transportation temperature 40 C 60 C...

Страница 11: ...ems in batches which is not exceeding 1 3 of the box volume at a time The next batch of goods can only be put in after the device is shut down for more than 12 hours 6 Try not to open the door during...

Страница 12: ...items are consistent with the environment in the freezer so as to avoid damage to the items and economic losses Due to the refrigeration inertia the display temperature and humidity of this product ar...

Страница 13: ...ement of products and different models the actual products may be different from the schematic drawings Please refer to the real objects The sketch is only used for the description of the functional p...

Страница 14: ...s in the display window in the upper chamber when the upper chamber door is not closed Door2 flashes in the display window in the lower chamber when the lower chamber door is not closed Fault indicato...

Страница 15: ...lt is not removed after 5 minutes of muting the buzzer will alarm again Press down again and the buzzer alarms After displaying the ambient temperature for 5s display of cabinet temperature and alarm...

Страница 16: ...gether to close the upper chamber The upper digital tube flashes and displays off In the normal operation mode first unlock the panel and press and hold the keys and together to close the lower chambe...

Страница 17: ...MAX1 and MAX2 parameters in the user menu as well as the MIN1 and MIN2 parameters display the maximum and minimum temperature values within the current time period Battery voltage check Press down th...

Страница 18: ...er maximum temperature in the current time period 4 MIN2 Lower minimum temperature in the current time period 5 CLR 0 1 1 0 Clear the record of the maximum and minimum temperature in the current time...

Страница 19: ...nth 14 r Check the current date day 15 S Check the current date hour 16 F Check the current date minute 17 Pt 0 240 minutes 20 Printing interval 18 tH1 20 0 50 0 32 Ambient temperature alarm value 19...

Страница 20: ...r opening alarm of the upper chamber door 2 Door opening alarm of the lower chamber PF Power failure alarm BL Low battery alarm EE Display board communication fault E1 Sensor 1 of the upper chamber al...

Страница 21: ...he top of the device to avoid deformation of the device under pressure Device Maintenance Cleaning and maintenance The interior of the freezer should be cleaned once a month Regular cleanings can keep...

Страница 22: ...osal in accordance with local regulations and ordinances non professionals should not disassemble or disassemble the equipment without authorization Scrapped equipment should be placed in designated a...

Страница 23: ...y Whether a part of the cabinet is in contact with external objects or walls During use it is normal for the actual noise to differ from the marked value due to the influence of loaded items ambient n...

Страница 24: ...aming door The lower chamber Foaming door Foamer Cyclopentane Shelf The upper chamber 6 drawers The lower chamber 6 drawers Cooling mode Direct cooling Shell plate Inner plate PCM painted steel plate...

Страница 25: ...erature sensor Upper chamber control sensor Lower chamber control sensor Upper chamber door switch Lower chamber door switch Alarm wiring block Condenser blower Normally closed Public Normally open Co...

Страница 26: ...he repairman who is not responsible for three guarantees Without warranty card and valid invoice The warranty card model s inconsistency with the model of the product to be repaired or the alteration...

Страница 27: ...s Two years The main components include compressor main control panel display control panel temperature controller evaporator condenser solenoid valve filter capillary fan motor Note if there are a se...

Отзывы: