Precautions for Use

  Before putting articles into the equipment, please make sure that the temperature 

in the storage room has reached the set temperature, and then put the articles in 
batches. The volume of the articles in the box should not exceed 1/3 of its volume 
at a time to prevent the temperature from rising too much.

  The temperature display value of the equipment is to show the temperature at 

the temperature sensor in the storage room. There is a certain gap between the 
displayed temperature and the actual temperature at the center of the equipment 
when the equipment just runs, but as the equipment enters a stable state, the 
displayed temperature will gradually approach the actual temperature.

  The equipment box body is provided with a detection hole so as to lead out the 

test line in the box during the test. After the test line is led out, it is necessary to 
plug the test hole with insulating material again, otherwise the temperature in 
the box may not reach the set temperature value, and condensation will appear 
around the outside of the through hole.

  Please use diluted neutral cleaner to clean the equipment, and do not use 

brushes, acid, gasoline, soap powder, polishing agent or hot water to clean 
the equipment, otherwise the painted surface and plastic rubber parts may be 
damaged. Attention: Do not use volatile solvents such as gasoline to clean plastic 
and rubber parts.

  When the device is not in use for a long time, cut off the power supply and turn off 

the battery.

  Each time you access the items, please try to reduce the time to open the door to 

avoid large fluctuations in temperature and humidity inside the box.

  After the door is opened, the temperature inside the box will rise sharply in a short 

time. This is normal, and the situation will be restored within 1 hour after the door 
is closed.

Содержание VS -25L Series

Страница 1: ...EN 61326 1 2013 Product implementation standards EN 61010 1 2010 A1 2019 EN 61010 2 011 2017 USER MANUAL MODEL VS 25L518 VS 30L518 VS 40L518 Biomedical...

Страница 2: ...r Use 06 4 Product Installation 07 5 Product Composition and Overview 11 6 Operating Instructions 12 7 Function Settings 15 8 Troubleshooting and Maintenance Services 20 9 Specifications 22 10 Packing...

Страница 3: ...rature data can be traced and the test data and the temperature curve inside the box can be output at any time Efficient Refrigeration Energy Saving Mute Vacc Safe is a world renowned compressor brand...

Страница 4: ...icians Be sure to place the freezer firmly on a solid and flat ground If the ground is not stable or the place is not suitable the freezer might be overturned or people will be injured If the power co...

Страница 5: ...en storing toxic harmful or radioactive materials please use a freezer in a safe area Improper use may cause harm to human health or the environment When the freezer is not used for a long time unplug...

Страница 6: ...ining water on the freezer If the object falls it may cause personal injuries and the outflow of water will reduce the insulation and cause leakage or electric shock Do not insert metal objects such a...

Страница 7: ...or personal injury Please do not use the door handle to lift or carry the equipment in case of injury to the freezer or personal injury The refrigeration circuit must not be damaged Do not use electri...

Страница 8: ...during the test After the test line is led out it is necessary to plug the test hole with insulating material again otherwise the temperature in the box may not reach the set temperature value and co...

Страница 9: ...it may cause electric leakage or electric shock Installation Site Do not install the device in a narrow and confined space Install it in a place that can ensure its normal access to avoid damage to st...

Страница 10: ...vice may not be able to operate normally Please improve the environment before using the device the device is working in intermittent operation Ambient storage and transportation temperature 40 C 60 C...

Страница 11: ...ems in batches which is not exceeding 1 3 of the box volume at a time The next batch of goods can only be put in after the device is shut down for more than 12 hours 6 Try not to open the door during...

Страница 12: ...items are consistent with the environment in the freezer so as to avoid damage to the items and economic losses Due to the refrigeration inertia the display temperature and humidity of this product ar...

Страница 13: ...ement of products and different models the actual products may be different from the schematic drawings Please refer to the real objects The sketch is only used for the description of the functional p...

Страница 14: ...s in the display window in the upper chamber when the upper chamber door is not closed Door2 flashes in the display window in the lower chamber when the lower chamber door is not closed Fault indicato...

Страница 15: ...lt is not removed after 5 minutes of muting the buzzer will alarm again Press down again and the buzzer alarms After displaying the ambient temperature for 5s display of cabinet temperature and alarm...

Страница 16: ...gether to close the upper chamber The upper digital tube flashes and displays off In the normal operation mode first unlock the panel and press and hold the keys and together to close the lower chambe...

Страница 17: ...MAX1 and MAX2 parameters in the user menu as well as the MIN1 and MIN2 parameters display the maximum and minimum temperature values within the current time period Battery voltage check Press down th...

Страница 18: ...er maximum temperature in the current time period 4 MIN2 Lower minimum temperature in the current time period 5 CLR 0 1 1 0 Clear the record of the maximum and minimum temperature in the current time...

Страница 19: ...nth 14 r Check the current date day 15 S Check the current date hour 16 F Check the current date minute 17 Pt 0 240 minutes 20 Printing interval 18 tH1 20 0 50 0 32 Ambient temperature alarm value 19...

Страница 20: ...r opening alarm of the upper chamber door 2 Door opening alarm of the lower chamber PF Power failure alarm BL Low battery alarm EE Display board communication fault E1 Sensor 1 of the upper chamber al...

Страница 21: ...he top of the device to avoid deformation of the device under pressure Device Maintenance Cleaning and maintenance The interior of the freezer should be cleaned once a month Regular cleanings can keep...

Страница 22: ...osal in accordance with local regulations and ordinances non professionals should not disassemble or disassemble the equipment without authorization Scrapped equipment should be placed in designated a...

Страница 23: ...y Whether a part of the cabinet is in contact with external objects or walls During use it is normal for the actual noise to differ from the marked value due to the influence of loaded items ambient n...

Страница 24: ...aming door The lower chamber Foaming door Foamer Cyclopentane Shelf The upper chamber 6 drawers The lower chamber 6 drawers Cooling mode Direct cooling Shell plate Inner plate PCM painted steel plate...

Страница 25: ...erature sensor Upper chamber control sensor Lower chamber control sensor Upper chamber door switch Lower chamber door switch Alarm wiring block Condenser blower Normally closed Public Normally open Co...

Страница 26: ...he repairman who is not responsible for three guarantees Without warranty card and valid invoice The warranty card model s inconsistency with the model of the product to be repaired or the alteration...

Страница 27: ...s Two years The main components include compressor main control panel display control panel temperature controller evaporator condenser solenoid valve filter capillary fan motor Note if there are a se...

Отзывы: