21
FPA
Temperature
Alarm
FPA Temperature is
out of acceptable
range.
1.
Disable TEC.
2.
Power off camera and let cool.
22
Camera
Alignment Error
Internal Error
Power Cycle Camera
23
Corr DL
Suspend
Correction Download
in Progress
No action required.
24
AGC High/Low
OPR Error
AGC High and Low
OPR settings are
invalid.
Correct OPR High/Low OPR settings.
25..31
(MSB)
Unused Unused
Unused
Table 8. Error value descriptions, causes and resolutions.
For text descriptions of the error bit fields use the command options ON and ALL.
Command
Response
ERROR?
Returns the 32-bit error register value only
ERROR? ON
Returns the error register value and its text description
ERROR? ALL
Returns the list of all error register values and their text descriptions
5.16.17.
Reboot Camera
Description
Execute the power-up initialization sequence of the command
processor. This will also clear the power-down detect flag.
Setting Type
N/A
Command REBOOT
Parameters none
Return Values
start-up banner
Range N/A
Type string
Example
REBOOT -- restart command processor