Run test
to ensure system integrity. Press
Start
button (if the start button is not on
the dashboard, use start on remote box). Allow test cycle to complete its 5 steps: Fill,
Isolate, Stabilize, test and Exhaust (see Figure 10). The result of this test should be
less than 0.3 mbar. The Optima vT will give a pass if within the tolerance.
See Figure 11.
Figure 10: Optima vT: Chamber Seal Test Run Screen
Figure 11: Optima vT: Chamber Seal Test Results Screen