5
Specifications
Air supply:
6 bar
Mains supply:
100v -240v
Weight:
Chamber Packed 40 kg, unpacked 35 kg/Optima Packed 42 kg, unpacked 32 kg
Chamber, lid and internal components
Anodised Aluminium.
Adjustment lock nut
Black acetyl
Hinges and all Screws
Stainless Steel
Latch and Striker plate
Hardened Stainless Steel
Pressure plate sheet:
Porvair
Sample Infills
Polypropylene
Lid seal
Silicone
O-rings
Nitrile
How it Works
The Qualipak-770 Package Test Fixture is designed to integrate with the Optima vT Leak
Tester to test the integrity of air tight packages/parts by performing a Vacuum Force Decay
test on the various packages/parts. The test part is loaded into the test fixture. The
operator selects the appropriate program for the part to be tested, then selects the start
button on the test fixture; the test automatically starts. During the test, the part under test
is subjected to a vacuum supplied by the Uson Leak Tester. The vacuum level is regulated
by the Optima Leak Tester, and during the test, the Qualipak-770 feeds back a signal
from a load cell to the Uson Leak Tester. The Leak Tester program then differentiates
between a good and bad part.