Built-in Data Logger
Once the calibration curve is sent to ETL3000/ETLONE, the reply is Y if the command has
been correctly received or N if there has been an error, in this case please try again.
The ETL can be provided with SENS3000 (old type sensors) or with the new SENS-IT sensors.
The calibration curves are different according to the type of sensors installed.
Examples of Calibration curves:
CO
@#L01;+0.279;+0.0002.0;** for SENS3000
@#L01;+0.0;+0.01718;+0.0;** for SEN-IT
O3
@#L02;-31.6;+5330.9;-0.598;** for SENS3000
@#L02;+0.0;+ 0.05012;+0.0;** for SEN-IT
NO2
@#L03;+1397.742;-164.415;** for SENS3000
@#L03;+0.0;+0.0523;+0.0;** for SEN-IT
C6H6
@#L04;-2.0003;+0.0062238;** for SENS3000
@#L04;+0.0;+0.00616;+0.0;** for SEN-IT
Temp(for MET BASIC)
@#L05;-40.0;+60.0;**Ö
RH (for MET BASIC)
@#L06;+0.0;+100.0;**ìf
Noise (curve 45-105 dB)
@#L07;+45.0;+60.0;**;š