Technical Description
2.3
Electric Connection
Equipment rating:
•
220 V - 50 Hz (mains cable is supplied approved by CEI regulation)
•
110V - 60 Hz (mains cable is supplied approved by CEI regulation)
•
Optionally : 12 V (cable is not supplied)
•
Absorption (without sensors and accessories):
220 V – 50 Hz 0,1 A
12 V
1,1 A
ETL is provided with a main switch having a protection 2A fuse (220 V) or 3A fuse (12V)
Before operating the instrument, check:
•
The correct rating to be used (check the power supply note on the label of ETL)
•
That a protective earth is installed
•
The integrity of the mains cable
All wiring must be in accordance with local norms and be carried out by experienced
personnel.
2.4
Connection of input signals
The microprocessor board of the ETL3000/ETLONE
can be set to receive 36 input 0-5V signals.
This option should be activated at Orion’s factory only by ORION’s personnel.
If required, please contact ORION’s service department.