Built-in Data Logger
Go on
‘
’ASCII SET UP’’
ASCII SENDING
Click on :
Echo typed characters locally
Set Line Delay ‘’ 0 ‘’ milliseconds
Set Character Delay ‘’ 0 ‘’ milliseconds
ASCII RECEIVING
Click on
Append line feed to incoming line ends
Wrap lines that exceed terminal width
Now click OK and OK.
Now the ‘’ETL’’ Hyperterminal session is created .
The commands recognised by this software version are:
1.
Command to memorize in FLASH memory the calibration curves of the gas sensors and of
the noise sensor
2.
Command to set date and time
3.
Command for data transmission of averages stored in SRAM
4.
Cancel the memory buffer
5.
System Restart, reconfiguration without cancelling the buffer and system pointers.
6.
Downolad meteo data.
6.1.1
Command to memorize the calibration curves
The set of command is :
@#L[Numch];+/-[xxxxxxxxxx].[xxxxxxxxxx];
+/-[xxxxxxxxxx].[xxxxxxxxxx];
+/-[xxxxxxxxxx].[xxxxxxxxxx];**[MSB crc][LSB crc]
[
Cr
]
where:
L
command identification
Numch=
channel number
+/-
firts parameter of the calibration curve
+/-
second parameter of the calibration curve
+/-
third parameter of the calibration curve
[MSB crc]
[LSB crc]
[
Cr
]
carriage return to end the set