10-3.
Error message list
-LoAd
Signal input range
Over by -5V
The input electric signal substantially exceeds the signal input
range. This occurs when an excessive load is applied to the
sensor, or when the input terminal becomes an open circuit due
to cable disconnection and so forth.
+LoAd
Signal input range
Over by +5V
oFL1
When below -99999
This occurs when a load greater than the assumed load is
applied to the sensor or when calibrating below the
measurement range. Check calibration, items to be measured
and so on.
oFL2
When over 99999
CALERR
Failed operation of capacity setting
This occurs when capacity setting cannot be configured
because UTM
Ⅱ
is not connected, UTM
Ⅱ
or cables are faulty
or cannot be detected, or the internal circuit is abnormal.
USBERR
Failure to acquire time
The internal circuit is abnormal.
DACERR
DAC calibration failure
The internal circuit is abnormal.
NODATA
There is no data to transfer to USB
This occurs when there is no recorded data, or when recorded
data is deleted.
USBERR
Communication failure with the
USB
The internal circuit is abnormal.
OVER/REC
Point data over
This occurs when attempting to record after the internally
saved point data reaches 1000.
* Displayed only while "REC" is pressed. When recording via
external signal, inputs are ignored and nothing is displayed.
USBERR/REC
USB record error
This occurs when the date and time could not be read when
recording data. When data is transfered to USB, the date and
time are displayed as --. Also this occurs when the date and
time are not set.
The internal circuit is abnormal.