7.1.7
Battery Test
Battery test is used to detect the battery capacity. Battery capacity is an important index of battery, it
reflect the using time and reliability issue of battery.
During the battery test, the voltage will decreasing with the increasing of discharge time, so it need to
set the cut-off voltage, the test will stop when it meets the cut-off voltage.
Parameter of Battery Test
In battery test, select discharge mode and set the load parameter and cut-off voltage for the mode.
The electronic load will automatically stop when the battery discharges to the cut-off voltage.
V
0
Battery Voltage
T
Cut-off Voltage
Time
Battery Test : CC Mode
Figure 7-1-7 Battery Test
Battery Test Mode
During the actual test, user can view the battery voltage, discharge current and discharged capacity at any
time.
Operation steps
1.
After the instrument is boot-up, press CH key to select the channel, yellow wireframe area is the
selected channel and it will display the prompt of CH1 or CH2 at top of the screen ( single
channel don’t need to choose.)
2. Press Mode key to enter the selection interface, use rotary knob or direction key to select battery
mode and then press Enter key to enter the set interface.
3. Rotate the rotary knob to select the parameter to adjust, press Enter key to change and press it again
to save the setting.
4. After the parameter is set, press ON key to enter list test interface (press ON1 to enter CH1, press ON2
to enter CH2).
5. Press ON key to operating, press ON key again to stop the operation.
Parameter
Description
Load Mode
Set discharge mode to CC/CR/CP
Load Size
Set the load value
Cut-off Voltage
Set the lower limit for cut-off discharge