User’s Manual Series DC Electronic Load
Instruments.uni-trend.com
19
/
33
7.1.5
Dynamic Test
series electronic load has dynamic current load mode. In dynamic mode, user can set
the two constant parameter, the electronic load can switch the two constant value by set the
operation mode. The following table is an example of the set interface in this mode.
[Dynamic] [Manual]
Operation mode
Continuous
Repeat Times
01000
Low value
1.000
A
Timing of low scale
100.0
ms
High value
5.000
A
Timing of high scale
100.0
ms
Rising Slop
0.100
A/us
Falling Slop
0.100 A/us
Parameter of Dynamic Test
Dynamic Test
Description
Operation mode
Set operation mode:continuous, pulse, reversal
Low value
Set the parameter value of low scale
Timing of low scale
Set the load time of low scale
High value
Set the parameter value of high scale
Timing of high scale
Set the load time of high scale
Rising Slop
Set the rising slop
Falling Slop
Set the falling slop
Repeat Times
Set the repeat times
Operation Description of Dynamic Mode
1.
Continuous Mode
The electronic load will automatically switch the two setting high/low scale until it reaches to the
repeat times and the test is over. Continous transient operation as shown in the following figure.
Figure 7-1-5-1 1 Continous Transient Operation
2.
Pulse Mode
The electronic load will running with the low value, and then the electronic load switches to the high
value when it receives a trigger signal for each time. After maintaining the set time, the electronic
load switches to the low value. The electronic load will only reverse one time when it receives a