User’s Manual Series
Instruments.uni-trend.com
29
/
40
Com mode
Baud rate
Port
USB – File name prefix - Prefix of storage file
USB – Time period per file – Automatic segnebtation of storage file
USB – delay – Time interval of data record
Figure9-1 <Function Setup> Page
Table 9-1-1 Function Setup
Item
Input Range
Default
Description
Com
ON, OFF
OFF
Turn on or off comparator
Beep
ON, OFF
OFF
Turn on or off comparator beep
Key Beep
ON, OFF
ON
Turn on or off key beep
Rate
Slow, fast
Slow
Fast: sampling period is 0.5s
Slow: sampling period is1s
Unit
℃
, K,
℉
℃
Temperature unit
AutoStart
ON, OFF
OFF
Auto-start after power outage, whether to
restore the measurement state after
reboot
COM BUS
RS232C, LAN
RS232C
Communication bus setting
COM MODE
Modbus, SCPI
SCPI
The
instruments
supports
two
communication protocols SCPI and Modbus
(RTU). In general, using SCPI to
communicate with computer and Modbus
to communicate with PLC.
Baud rate
4800, 9600, 19200, 38400,
9600
Baud rate of serial port bus