Get ready to start a measuring campaign
33/92
4.3 Attach the sampling ring
4.3.1 Auger the Tensiometer holes
Rotate the auger while pulling it out of the sample.
1. Take the soil sample out of
the saturation dish. Place the
auger positioning tool on the
sampling ring as shown in the
picture.
Fig. 31
the deeper hole is. Then you
know the correct position when
placing the sample on the
sensor unit.
Fig. 32
3. The sample now is ready for
attaching the sensor unit.
2. Insert the auger into each
opening and drill a hole. Drill as
far as the auger will go. Rotate
the auger while pulling it out of
the sample. Then, you will have
2 holes in the sample, each with
the proper depth.
With a pen make a mark on the
side of the sampling ring where
Fig. 32
Fig. 33