Set-up the HYPROP
43/92
Soil volume
Volume of the soil sample in the sampling ring excluding the
Tensiometer shafts’ volume.
Soil column height
Height of the soil column in mm (height of sampling ring).
Depth lower tens
Protruding length of the lower shaft.
Depth upper tens
Protruding length of the upper shaft.
Data logger
Interval
the logging interval of the internal data logger
Overwrite old values
Overwrites old values (if you select „on“) if the memory is full
Sensor measuring
Continuous measuring
Activate the quick updating of readings to receive the HYPROP
readings instantly, for example during a refilling procedure.
Measurements are taken in intervals of 50 ms. Note the rise in power
consumption and that the reaction to serial commands might be
slowed down. The setting "Measuring interval“ is deactivated during
this mode..
Measuring interval
This is the standard interval in which sensor measurements are
refreshed and available on the analogue lines.
Enable filter
Activate the anti-flicker-filter. This avoids that the digit continuously
jumps up and down. When activated the resolution is reduced for
one digit.