11/92
3 Product
description
3.1 System components
A measuring system can include one or several HYPROP
assemblies (max. 20). A HYPROP assembly consists of a sensor
unit and a sampling ring with a soil sample which is placed on each
sensor unit. Sensor units are linked to a PC via the serial
tensioLINK
®
bus.
In intervals each sensor unit with sampling ring is weighed on a
laboratory scale. The scale must have either a RS232 or USB
interface, and the scale type must be implemented in the software.
3.2 Sensor unit
3.2.1 Main body
The electronic components and pressure transducers are
incorporated in the main body of the sensor unit. The sensor unit is
splash water proof (IP65) and can be cleaned with water as long as
the plug cover is closed.
Fig 2: HYPROP assembly