Get ready to start a measuring campaign
20/92
4.2.2 Introduction
Before starting and after every completed measuring campaign (the
sample is dried out) the Tensiometers need to be filled or refilled
bubble free with deionised and degassed water.
Keep the ceramic moist when not in use by covering them with the
water filled rubber cap.
Ceramic tip: Do not touch the tip with your fingers. Grease, sweat
or soap residues will influence the ceramic's hydrophilic
performance.
The HYPROP-service kit or a vacuum system and a PC/Laptop with
tensioVIEW
®
software are required for filling or refilling.
4.2.3 Refilling with the HYPROP service kit
This chapter describes the refilling using the tools of the HYPROP
service kit.
The procedure requires 5 steps, which will be discussed in detail
later in the chapter:
1. Check the Tensiometer
2. Degas ceramic cup and shaft
3. Degas the sensor unit
4. Reassemble
5. Check the result