4.6 Oil mist trap (Optional)
The oil mist trap can be mounted to trap the oil mist discharged from the Pump.
Pressure at the inlet
10000Pa or less
10000Pa or more
VD30C/VD40C
TM201
TM401
VD60C/VD90C
TM401
TM-2
(With an adapter)
For more information, please see the instructions OIL MIST TRAPS.
Fig. 18 Installation example VD90C
WARNING
Limit value of the pressure inside the Pump is 0.03MPaG (0.3kg/cm2G) (Gauge
pressure). If the pressure is raised, filter element please exchange.
We recommend you to install a Pressure monitor.
To use the Oil mist trap, be cautious not to have clog of the filter in the trap. Too
much clog would prevent the exhaust gas from passing through the filter, raise
the pressure inside the Pump unit (include the oil mist trap) and might result in
breaking it.