13
3.6.
25B
View when the Bottom Panel in Controller is opened
Fig. 3-6 View when bottom panel in controller is opened
Name (Inscription)
Function
①
Internal DIP switch
※2)
(DS1)
Switch for settings.
DIP switch
No.
ON
OFF
1
Mantissa part 2
digits display
Mantissa part 3
digits display
2
※1)
Pressure
protection function
9.9E-3
Pa
(7.4
×
10
-5
Torr)
ressure protection
function
10 Pa
(7.4
×
10
-2
Torr)
3
Not used
Not used
4
Not used
Not used
5
Used in system setting.
The setting cannot be changed.
6
7
8
②
External input switching
jumper
(HP3)
Jumper for selecting whether the internal power supply
(5 V) or the external power supply is to be used to pull
up the external input photocoupler. See P. 63
③
External output switching
jumper
(HP4)
Jumper for selecting whether the common line of the
external output open collector (emitter common) is to be
connected to pin 5 of the external I/O connector or to the
internal ground. See page 59
※1)
Use it only in GI-M2. Turn it on at the time of tungsten filament use.
※
2
)
When the factory shipment, it is set entirely by OFF.
Control boad
F
ro
nt
p
anel
R
ea
r p
anel
1
ON
2
3
4
8
5
7
6
HP3
EXT
INT
HP4
ISO
GND