7
Power
requirements
AC100±10V
50/60Hz
Power
consumption
60VA
120VA
65VA
Fuse
BS.SEMKO EWM 3.15A (Fuji Terminal Industry) (2 pc)
Outside
dimensions
240W
×
380D
×
99H (mm)
Weight
5.6
kg
5.7
kg
5.8kg
※
1
)
Pressure display
can set that mantissa part 2 digits, exponent part 2 digits digital display
(
□
.
□□
×
10
□□
).
※
2
)
10
µ
A and 1.0 mA are automatically changed over depending on pressure,
1.0 mA is available when the pressure measurement value is 0.79
×
10
-2
Pa or less
※
3
)
Emission current of 5.0 mA is available under a pressure of 6.6
×
10
-4
Pa or less.
※
4
)
When use a tungsten filament, it is necessary to do pressure protection of a set point of
functioning in 9.9
×
10
-3
Pa
※
5
)
Available under a pressure of 1.3
×
10
-4
Pa or less.
2.2. Standard Accessories
External input/output connector
57-30500: DDK or equivalent
1 pc.
Power cable
125 V, 7 A, 3-pin cable, 3 m long
1 pc.
Quick manual
Ordinary paper
1pc
This gauge does not come with the RS-232C communication cable.
Use a commercially available RS-232C cable (cross type).
2.3. Options
Sensor head
Refer to P.86
~
for the each sensor head details.
Refer to P.86
~
for the demontions details.
Sensor head cable
Refer to P.89
~
for the details.
Inspection record
Calibration certificate
Calibration certificate, JCSS Calibration certificate
Содержание GI-M2