background image

 

-22- 

6.

Maintenance, Inspection and Repair 

6.1 

Maintenance 

Check the following during operation at least once every three days. 

(1)

 

Is the amount of the pump oil in the range of the red circle of the oil level gauge? 

  (2)  Does not the vacuum pump oil discolor? 

  (3)  Is not an abnormal sound done? 

(4)  Is not abnormality found in the value of the current of the motor? 

(5)   Is not there the oil leak? 

If there is any problem, take proper measures in accordance with “6.4 Trouble check list.” 

 

6.2 

Periodic inspection  

The items to be checked should be changed as necessary depending on the environment 

where the pump is used. However, always check the following in order to prevent a 

malfunction and to lengthen the service life of the pump.

 

 

 

 

Pulled out the power plug before starting inspection and do not turn it on while 

inspection is in progress. Doing so will result in injury. 

The pump is very hot immediately after it is stopped. Wait for 10 minutes until 

the pump has cooled down completely and then start inspection. There is a risk of 

burns.

 

 

(1)  Periodic replacement of the pump oil 

The pump oil deteriorates with operation. Check the viscosity and level of 

contamination of the pump oil with the oil level gauge, and replace the pump oil 

in good time.  

If the pump oil is replaced periodically, the deterioration of the pump oil is 

minimized and the service life of the pump is lengthened. 

If operation is continued with a lot of moisture mixed with the pump oil, the 

ultimate pressure will not reach the standard value, the movement at the section 

where the mechanical friction is generated becomes slow, and the pump finally 

becomes damaged. Replace the pump oil in accordance with “6.3 Replacement of the 

pump oil.” 

          Caution 

Содержание GHD-031A

Страница 1: ...s manual thoroughly to ensure safe and effective use of the equipment Keep this manual in a place where it can be referred to at any time and look after it carefully The contents of this instruction m...

Страница 2: ...igh temperature 3 1 2 Safety Data Sheet SDS 4 2 Outline of the Pump 5 2 1 Specification 5 2 2 Dimensional drawing 6 3 Measures before it starts 7 3 1 Securing of exhaust vent 7 3 2 Installation 9 3 3...

Страница 3: ...performance 19 5 1 Ultimate pressure 19 5 2 Pumping speed 20 5 3 Power requirement 21 6 Maintenance Inspection and Repair 22 6 1 Maintenance 22 6 2 Periodic inspection 22 6 3 Replacement of the pump...

Страница 4: ...Pumping speed curve 20 Figure7 Characteristics of motor power 21 Figure8 Lubrication of the oil sealed rotary vacuum pump 27 Figure9 Disassembly drawing of GHD 031 oil sealed rotary vacuum pump 33 Tab...

Страница 5: ...y pump specifications and method of operation can operate this pump Read the manual thoroughly Please completely read a manual well to use this product definitely Please read the section on introducti...

Страница 6: ...at the time of the unpacking Although the pump is delivered with great care check the following after unpacking 1 Confirm whether it is the model of the pump of the order 2 The specified accessories h...

Страница 7: ...tal accident or serious injury When mishandled there is a possibility of the operator suffering a fatal accident or serious injury When mishandled there is a possibility of the operator suffering an i...

Страница 8: ...y before installing electrical wiring or performing any electrical work on the pump Failure to do so may result in electric shock The inlet port of the pump The outlet port of the pump Various label p...

Страница 9: ...05 Table for various labels No Label name Label Rotation Burn caution Electric shock caution Installation caution Transportation caution Pump...

Страница 10: ...06 No Label name Label Inlet Position High temperature caution Outlet Pump oil R 2 Passed Warning explosion...

Страница 11: ...ist to make it safe Ask an authorized waste disposal specialist to carry out disposal This pump is for dry air or dry nitrogen absorption of the inert gas This pump cannot absorb poisonousness and fla...

Страница 12: ...outlet side When an oil mist trap is adopted replace or clean it so that it will not block the passage of gas There is a risk of explosion Never block the outlet or operate the pump with equipment mo...

Страница 13: ...e be sure to protect and cool surface of vacuum pump and away from human body Use this pump as built in type Since the surface temperature is hot touching the surface accidentally may result in burns...

Страница 14: ...materials vacuum pump oil other than those mentioned in this instruction manual SDS is submitted as reference information to ensure safe handling of hazardous and harmful materials Personnel handling...

Страница 15: ...80 110V 1 89 100V 1 71 120V 0 94 200V 0 84 240V 1 02 200V 1 03 240V Motor speed r min 2500 100V 2600 110V 3100 100V 3300 120V 2400 200V 2600 240V 2900 200V 3100 240V Standard oil R 2 Oil Oil amount m...

Страница 16: ...6 2 2 Dimensional drawing Figure1 Dimensional drawing of GHD 031 oil sealed rotary vacuum pump...

Страница 17: ...curing of exhaust vent Please remove the oil cap for transportation without fail before starting the pump and exchange it for the attached exhaust entrance cap At the unpacking Detaching of oil cap In...

Страница 18: ...arting the pump and exchange it for the attached exhaust entrance cap There is a risk of explosion Never block the outlet or operate the pump with equipment mounted at the outlet side which blocks the...

Страница 19: ...temperature relative humidity 7 C 40 C 85 RH or less Height above sea level during storage and installation 1000 m or less Minimum required distances from the wall 100mm or more Other conditions for s...

Страница 20: ...n mechanism of the oil is taken in in this pump As of a pump stop it is effective for the prevention of the countercurrent of the oil It is not countercurrent prevention mechanism of the gas The count...

Страница 21: ...than oil countercurrent quantity to the vacuum room side approximately 300mL at the time of a pump stop in a short time between a stop valve A which I flow backward and figure 2 shows to by all means...

Страница 22: ...tric shock occurs 1 The pump rotates in the clockwise direction as seen from the front of the pump level gauge side Figure3 Electric wiring diagrams 2 An inlet is suitable for standards of EN60320 Fig...

Страница 23: ...s a model GHD 031A 100 120V time and a model The choice standards of the power supply cord are more different in GHD 031B 200 240V The power cable selections standards are shown in a table 2 Do not fa...

Страница 24: ...e with the procedure as follows Let one end of the unplug preventive hardware Catch the hole to the side of the inlet in the interior While pushing the unplug preventive hardware at the other end let...

Страница 25: ...ut the obstacle on practical use and driving the voltage change and the frequency veering in area B in the main rated value without the obstacle on practical use However operation with it is convenien...

Страница 26: ...p an oil level gauge jumps out In addition This causes the pomp rock Please change pumps oil 4 2 Start of operation To start operation close leak valve B open vacuum valve A to the inlet port and turn...

Страница 27: ...uous operation may become possible After the pump has warmed up close leak valve B and return to ordinary operation see Figure2 4 5 Thermal protector An overload protector Auto reset thermal protector...

Страница 28: ...rt from the outlet port of the pump and install an oil mist trap instead The oil mist trap not only prevents oil mist generation but also reduces exhaust noise For details refer to the instruction man...

Страница 29: ...port Also the actual ultimate pressure of the vacuum device becomes higher than that noted in t he catalogue for the following reasons The vacuum gauge is installed at a distance from the pump and th...

Страница 30: ...mping speed usually reaches the maximum at a high pressure range and it gradually decreases as the pressure reduces The nominal pumping speed of this pump is the maximum pumping speed when dry air is...

Страница 31: ...required to compress the air compression work and reaches a maximum at an inlet evacuation pressure of around 2 7 104 to 4 104 Pa If the inlet evacuation pressure has reduced to 13 3 Pa or less the co...

Страница 32: ...starting inspection and do not turn it on while inspection is in progress Doing so will result in injury The pump is very hot immediately after it is stopped Wait for 10 minutes until the pump has coo...

Страница 33: ...Check the nuts and bolts for looseness 5 Inspection of the oil mist trap When using the oil mist trap in replacement of the standard outlet port pay attention to the clogging of the filter in the oil...

Страница 34: ...perature higher than the room temperature by 40K or more is abnormal Check the cause of an overload If not clear contact us Once week Oil leakage Oil leakage from the plugs Replace seals or contact us...

Страница 35: ...due to evacuated gas but also by the changes in the properties of the pump oil itself depending on the operation time Periodic replacement in accordance with Table 4 showing an oil replacement guide i...

Страница 36: ...essure even after the pump oil is replaced In such a case overhaul the pump 7 Remove the oil inlet plug from the oil inlet port and add the pump oil which has been delivered together with the pump or...

Страница 37: ...27 Figure8 Lubrication of the oil sealed rotary vacuum pump Do not operate the pump without adding pump oil If it is operated in oil less condition the pump will be damaged Note...

Страница 38: ...the cylinder and rotor Moisture or solvents were sucked in forming rust inside the pump Overhaul replace the cylinder and rotor Reaction product accumulated in the pump when the pump stops after exha...

Страница 39: ...nt of oil has not been added Add the specified amount of oil The oil has deteriorated Replace the oil Leakage occurs from the pipe connected to the pump Locate the leakage with a leakage detector and...

Страница 40: ...ount is not sufficient the cooling effect of the pump will be reduced Add the specified amount of oil The temperature of the evacuated gas is high Mount cooling equipment such as a gas cooler at the i...

Страница 41: ...he pump In case a harmful toxic gas has been exhausted by accident ask a specialist for waste disposal Not only the pump itself but also the pump oil becomes toxic For the disposal of pump oil follow...

Страница 42: ...O ring_S 60 1 39 O ring Casing 1 40 O ring_P 5 1 41 O ring_P 8 1 42 O ring_P 9 2 43 O ring_P 12 5 1 44 O ring_P 18 1 45 O ring_G 50 1 46 Oil seal _VC 10 20 4 1 47 Oil seal _SC 11 25 7 1 81 O ring_SS04...

Страница 43: ...33 8 2 Disassembly drawing Figure9 Disassembly drawing of GHD 031 oil sealed rotary vacuum pump...

Страница 44: ...odifications or repairs carried out by a party other than the manufacturer or by a service company not approved by the manufacturer e Malfunctions caused by noise electric disturbance f Malfunctions t...

Страница 45: ...fety and Health Law designates particular substances as the materials to be notified 2 Usage Status Operation Method Approx hours per day years and months Continuous Operation Intermittent Operation U...

Отзывы: