[Checksum OFF]
This instrument receives the checksum value of the data on the host side, but does not collate
it with command data. Whatever the checksum data is, it is returned according to it if only the
command is correct. (The checksum value of the data on the host side can be communicated
as fixed to “00” or other.)
[Checksum ON]
This instrument receives the checksum value of the data on the host side and collates it with
command data. If any error occurs in sending/receiving data as a result of collation, it is
returned as “n” and no action is taken against the command. If the result of collation agrees
with the command data, it is returned accordingly.
22.2.5.
Setting the terminator
If the total length of the connected cable is 15 m or more or if the frequency of communication
errors is high, it is recommended to install a terminator in the terminating device. Turn it ON on
the terminating device of the transmission line.
Display
Detail
“o”
Termination resistance
“F”
No termination resistance
22.2.6.
Setting method
Press the PROG key to change over the setting mode. Each press on the program key changes
the display as follows.