51
14.3.
Setting Operation
14.3.1.
Setting Overview
The following settings are available for turning on the filament or degassing and for changing
over the filament 1/2.
Setting
Overview
RS485 communication mode
Only RS485 communication is available.
Front panel operation mode
Only front panel operation is available.
External I/O operation mode
Only external I/O operation is available
14.3.2.
RS485 communication operation setting “L1”, “L2”
Selects if filament OFF, degassing ON/OFF or filament 1/2 changeover is to be operated
through RS485 communication. Refer to Section 22 for more information about the detailed
setting of RS485.
Display
Details
Remarks
“L1”
Operation on the front panel and external I/O
RS485: Pressure reading only
“L2”
Only operation through RS485 communication
14.3.3.
Front panel, external I/O operation setting
14.3.3.1.
Filament operation setting “Fi”
Selects whether filament OFF is to be operated on the front panel or with external I/O.
This function will be invalidated if it is set at RS485 communication.
Display
Detail
“F”
Operation on the front panel only
“o”
Operation of external I/O only
14.3.3.2.
Degas operation setting “dE”
Selects whether degassing ON/OFF is to be operated on the front panel or with external I/O.
This function will be invalidated if RS485 communication is set.
Display
Detail
“F”
Operation on front panel only
“o”
Operation of external I/O only
14.3.3.3.
Filament changeover setting “FL”
Setting only filament1, if you used ST2
ST2 has only one filament. Setting only filament1.
Selects whether selection of filament 1/2 is to be operated on the front panel or with external
I/O.
This function will be invalidated if it is set at RS485 communication.
Also the filament set on the front panel will be invalidated when operation is changed over to
external I/O.
Display
Detail
“0”
Operation of external I/O only
“1”
Setting of filament 1
“2”
Setting of filament 2
14.3.4.
Operation setting method
Press the program key to change over the set mode. Each press on the program key changes
over the display as follows. If Filament is ON (local or remote), this unit cannot do filament
setting, degas Setting, and filament 1/2 setting.