![Ulvac G-Tran ISG1 Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/ulvac/g-tran-isg1/g-tran-isg1_instruction-manual_837695037.webp)
25
8.2.
Input signal
Contact capacity
Use a capacity for externally installed contacts greater than the input
power supply voltage and 30 VDC or higher.
Contact leak current
Be aware of contact leak current. If a current of 0.08mA or higher flows
between the input signal pin to the GND terminal, that may be treated as
signal input.
Actuated when shorted to GND.
Fig.8-2 Input signal
8.2.1.
RS485 communication
Refer to Section 22 for RS485 communication.
5
:
FIL ON/OFF
6
:
FIL 2 / 1
13
:
DEGAS ON/OFF
9,15
:
GND
Display
Internal 5V
power supply
Internal 24V
power supply