62
16.
Multi-Ionization Gauge SH2/ST2(SWU Combination mode)
The support model becomes manufacturing numbers after 01000
If manufacturing number is before 01000, this unit may not output
setpoints.
We recommended a connection with the standard type
When connecting to the SH2-2/ST2-2 serial communication type, FIL ON,
etc., cannot be performed with the sensor. Depending on the cable
connection, there is a risk of damage to the SH2-2 /ST2-2multi-ionization
gauge.
Sensor and SH2/ST2 connection cable
The cable to connect the sensor and the SH2/ST2 has a maximum length of
20 m for the 24AWG. To use the sensor with a cable longer than 20 m,
directly connect the SH2 to a power supply or increase the cable diameter.
This section describes the operation of the SH2/ST2 (SWU Combination mode) ionization
vacuum gauge. Before reading this section, read the standard operating procedure in the foregoing
sections.
16.1.
Sensor Unit Setting: “SAU”
Turn on the power to this unit and make sure that “SAU” lights for several seconds.
To display others, press the
key several times within 3 seconds after turning on power,
change the display to “SAU” and make it definite with the
key.
For more information about the sensor unit setting, refer to Section 6.
16.2.
Front Panel
16.2.1.
Pressure display
Status
Pressure indicator
LED
In normal measurement
5.0
×
10
-8
Pa to 1.0
×
10
+5
Pa
Below measurement pressure range
0.0
×
10
-8
Pa
Above measurement pressure range
F.F
×
10
+F
Pa
When error of B-A gauge
When error signal is input
Pressure indication more than
1.0
×
10
-1
Pa
Err LED
lights
When error of Pirani gauge
When error signal is input
Err
(Setpoints off)
Err LED
lights
When error of Pirani gauge
When error signal is not input
Err
(Setpoints off)
B-A gauge Emission Valid OFF
*1
Display is pressure
(Setpoints off)
Err LED
blinking
When B-A gauge degassing is ON
Pressure indication
DEG is lit
B-A gauge FIL OFF
Pressure indication more than
1.0
×
10
-1
Pa
*1: However, the B-A gauge error is cleared by turning FIL off (Forced FIL OFF).
*2: When a CALCULATION function is used, it displays to 10
-11
at the minimum.
10
-10
is “A (A of capital letter)”, 10
-11
is “b (B of small letter)”.Please note “b (B of small
letter)” to make a mistake as '6' of the figure.
16.2.2.
Description of front panel keys
Inscription
Function
Remarks
PROG Enters into program mode
Refer to Section 5.
ZERO Not used
DEG
Degassing ON/OFF operation
Cannot be operated when RS485 and external I/O
are set
FIL
Filament OFF operation
Cannot be operated when RS485 and external I/O
are set