Maintenance and Repair 

(9)    Dispose in accordance with legislation for disposal and cleaning of waste products, handle as 

industrial waste, and do not incinerate. 

 

 

Toxic fluorine gas is generated by incineration of fluorine-based plastics. 

(10)  If the pump ceases operation, turn power OFF (set switch to O) immediately to prevent 

accidents, remove the power cord from the wall outlet, and contact your dealer or the 
manufacturer for inspection and repair. 

(11)  Leave the pump for at least 30 minutes until it has cooled, and begin operation again. Touching 

the pump immediately after it has stopped may result in burns.

 

 
 

 

 

Installation 

(1)  The pump may malfunction if it is subjected to shocks or 

tipped over on its side. 

(2)  Do not hold or push the tube at the top of the pump (see 

below). Damage to the tube may affect performance of 
the pump. 

 
 

Applications 

(3)  This pump is not designed to be corrosion-proof. Use it only with clean air at normal 

temperature, or with gases of equivalent characteristics. 

(4)    This pump is designed for general corrosion resistance, however it is not resistant to molten 

alkali metals such as molten sodium, to fluorine at high temperatures, and to some oxides of 
fluorine. 

(5)  Corrosion-resistant plastic is used in the external covering of the DTC-22, however it is not 

resistant to all chemicals. 

 

Ensure that the following chemicals do not come in contact with the pump. Any chemical, 
including the following, which comes into contact with the pump should be wiped off 
immediately. 

   

•    Acetone    • Ethyl ether    • Ethyl acetate    • Animal fats 

 

 

 

!

Caution

 

 

!

Note 

 

06

 

Содержание DTC-22

Страница 1: ...to Users Please read this manual thoroughly to ensure safe and effective use of the equipment Keep this manual in a safe place Due to periodic improvements in performance the equipment described in t...

Страница 2: ...hecking Operation After Installation 4 3 5 Piping 5 3 6 Storage 5 4 Cautions for Operation 5 4 1 Cautions for Operation 5 4 2 Operation of the Thermal Protection Relay 5 4 3 Precautions when starting...

Страница 3: ...sions 3 Fig 3 1 Example of Piping Used When Evacuating a Vessel 5 Table 1 1 DTC 22 Product Specifications 2 Table 1 2 DTC 22K Product Specifications 2 Table 6 1 Consumables List 7 Table 6 2 Locations...

Страница 4: ...quipment correctly Read the section on Safe Use particularly closely Keep This Manual in a Safe Place After reading this manual be sure to keep it in a safe place which is readily accessible to others...

Страница 5: ...ached to power cord x 1 It is attached only to 100V in the motor ratings voltage 3 Is the pump damaged in any way 4 Are any external screws or inlet and outlet pipes loose Are any components missing C...

Страница 6: ...y to the operator Incorrect handling of the equipment may result in death or serious injury to the operator Incorrect handling of the equipment may result in light or medium injuries to the operator o...

Страница 7: ...cking or repairing the pump Failure to do so may result in electric shock or the pump suddenly starting and causing injury 5 Ensure that the relevant wiring is in accordance with technical standards f...

Страница 8: ...rage installation and operation 1 Ambient temperature of 0 40 C and maximum relative humidity of 85 during operation 2 Other conditions for storage and operation a Level floor of sufficient strength b...

Страница 9: ...shocks or tipped over on its side 2 Do not hold or push the tube at the top of the pump see below Damage to the tube may affect performance of the pump Applications 3 This pump is not designed to be c...

Страница 10: ...of 40 C Use at high ambient temperatures will dramatically reduce the life of the pump 10 Back pressure at the outlet while the pump is starting may overload the motor 11 The thermal protection relay...

Страница 11: ...is pump is designed for general corrosion resistance however it is not resistant to molten alkali metals such as molten sodium to fluorine at high temperatures and to some oxides of fluorine 4 Corrosi...

Страница 12: ...istance to the tip of the intake or exhaust pipe 142mm W 272mm L 202mm H Voltage Category Pollution Degree Table 1 2 Product Specifications Model DTC 22K Discharge rate 50Hz 20L min 60Hz 24L min Press...

Страница 13: ...if the motor temperature rises due to a pump fault which prevents rotation or if load becomes excessive 2 It is recommended that additional protective devices eg earth leakage breaker motor breaker b...

Страница 14: ...l dust and humidity This location should be selected in consideration of ease of installation and removal inspection and cleaning Particular attention should be paid to ambient temperature when fittin...

Страница 15: ...en evacuating a vessel ensure that a shut off valve is placed between the pump inlet pipe and the vessel see Fig 3 1 Fig 3 1 Example of Piping Used When Evacuating a Vessel 3 6 Storage Turn the switch...

Страница 16: ...ure achieved as employed in the catalogue and in this manual is defined as the minimum pressure obtained by the pump without introduction of gas from the pump inlet ie the no load condition Note that...

Страница 17: ...lean in accordance with the Replacement and Cleaning Guide on the following page Refer to 6 4 Replacing and Cleaning Consumables for procedures Request replacement by the manufacturer s service divisi...

Страница 18: ...ing and Cleaning Consumables See Caution 11 P06 Always use a mask and gloves when replacing components to prevent ingestion of microscopic particles resulting from wear of components Use the following...

Страница 19: ...he pump head as shown in Photo 4 Note Gripping or bending the connecting tube will damage it and reduce performance of the pump Note After removing the pump head lay it on a soft cloth Note There is a...

Страница 20: ...ump head cover in place 8 Clean the tapped portion of the connecting rod by wiping with a cloth impregnated with solvent 9 Apply a small amount of vacuum grease to the threads on the new diaphragm and...

Страница 21: ...opposite pairs to a torque of 11Nm Note Confirm the thing that the pump head spacer has been inserted in the interior Photo 15 and 16 Photo 15 Photo 16 15 Position the top panel and fix it in place wi...

Страница 22: ...5 Fit the new inlet and outlet valves Note Ensure that the valves are fitted correctly as shown in Photos 18 and 19 6 Fit the valve retainers to the pump heads Note Ensure that the inlet and outlet v...

Страница 23: ...s 1 2 2 Remove the O rings See Photos 20 21 22 3 Clean the inlet and outlet pipes pump heads and valve retainers with solvent 4 Lubricate the entire surface of each O ring with a light coating of vacu...

Страница 24: ...amaged components 6 4 Pressure does not diminish 1 Pump is too small for capacity of vacuum vessel 1 Select another pump 2 Pressure measurement is incorrect 2 Measure the pressure correctly 5 1 3 Vacu...

Страница 25: ...e and inspection etc d Malfunctions caused by modifications or repairs carried out by a party other than the manufacturer or by a service company not approved by the manufacturer e Malfunctions caused...

Страница 26: ...Industrial Safety and Health Law designates particular substances as the materials to be notified 2 Usage Status Operation Method Approx hours per day years and months Continuous Operation Intermitte...

Страница 27: ...6 6397 2286 ULVAC KIKO Inc https ulvac kiko com en Please contact us for products Service Base or other Inquiries from here ULVAC Inc Components Division 2500 Hagisono Chigasaki Kanagawa 253 8543 Japa...

Отзывы: