background image

 

 

11 

7. CAUTIONS IN OPERATION 

 

For safe and trouble-free operation, please pay attentions for the following items in operating pump. 

  7.1 Cautions in Operation 

(1) 

This  pump  is  designed  exclusively  for  vacuum  pumping.    So  backing  pressure  of  the  pump  is 

needed to keep lower than 30kPa (Gauge pressure) by a reasonable means. 

(2) 

Use the pump at an ambient temperature of 7~40°C. 

(3) 

Use  the  pump  in  an  environment  where  it  is  not  likely  that  dust,  dirt,  etc.  are  admitted  into  the 

pump. 

(4) 

This  pump  is  not  designed  explosion-proof.    So  do  not  use  it  in  the  vicinity  of  inflammable 

solvent or the like, which can be very dangerous. 

(5) 

This pump is only for dry air.    When a large amount of vapour enters, it causes the breakdown. 

(6) 

Start the pump with the suction side and exhaust side under atmospheric pressure.    Otherwise, 

it is not so easy to start the pump. 

(7) 

If the pump is stopped under reduced pressure and is started immediately  after  that,  the  pump 

may not restart.    To restart it, vent the pump to atmospheric pressure. 

 

If the pump is used for other applications, consult with us. 

7.2 Startup in Cold Season 

 

In  a  cold  season,  it  may  sometimes  be  difficult  to  start  the  pump  due  to  hardened  bearing  grease, 

diaphragm or others.    In that event, proceed as follows. 

(1) 

Vent the suction port to atmospheric pressure and turn ON/OFF the power switch until the pump 

starts. 

(2) 

Operate  the  pump  for  several  minutes  with  the  suction  port  vented  to  atmospheric  pressure  to 

warm up the pump. 

(3) 

Operate the pump as usual after the pump is warmed up. 

7.3 Thermal Protector 

(1) 

This pump incorporates a auto reset type thermal protector.    It automatically cuts off the motor 

power circuit to prevent the motor from being damaged by burning when the pump stops rotating 

due  to  trouble  or  when  the  pump  is  energized  with  an  overcurrent  due  to  overload  during 

operation. 

(2) 

If the thermal protector is activated, turn OFF the power switch and contact our representative or 

ULVAC KIKO.    Never touch the motor, which is still very hot. 

(3) 

Restart operation after eliminating the cause of trouble and verifying that the motor temperature 

has lowered. 

 

  Characteristics of the t

hermal protector

 

Operation temperature 

135±5

 

Reset temperature 

85±10

 

 
 

Содержание DA-20D Series

Страница 1: ...100V B 115V C 220V ULVAC KIKO Inc Request to Users Please read this manual thoroughly to ensure safe and effective use of the equipment Keep this manual in a safe place Due to periodic improvements in...

Страница 2: ...9 Compressors and vacuum pumps Safety requirements Part2 Vacuum pumps IEC EN 61010 1 2010 A1 2019 Safety requirement for electrical equipment for measurement control and laboratory use Part1 General r...

Страница 3: ...elated is in conformity with the following standards BS EN 1012 2 1996 A1 2009 Compressors and vacuum pumps Safety requirements Part2 Vacuum pumps BS IEC EN 61010 1 2010 A1 2019 Safety requirement for...

Страница 4: ...ON 10 7 CAUTIONS IN OPERATION 11 7 1 Cautions in Operation 11 7 2 Startup in Cold Season 11 7 3 Thermal Protector 11 7 4 Fluctuations in the power voltage and frequency 12 8 Maintenance Inspection and...

Страница 5: ...quipment correctly Read the section on Safe Use particularly closely Keep This Manual in a Safe Place After reading this manual be sure to keep it in a safe place which is readily accessible to others...

Страница 6: ...outlet caps fitted to inlet and outlet 2 3 Is the pump damaged in any way 4 Are any external screws or inlet and outlet pipes loose Are any components missing Contact your agent or the sales division...

Страница 7: ...uipment may result in death or serious injury to the operator Incorrect handling of the equipment may result in light or medium injuries to the operator or damage to the equipment Incorrect handling o...

Страница 8: ...ir the pump always note the gas which the pump has been used with on the Usage Check Sheet Note that if it has been used to discharge toxic gas for any reason it will be contaminated Please be aware t...

Страница 9: ...ectric shock 15 Touching electrical wiring etc while inserting the power plug may result in electric shock 16 The electric motor fitted to this pump incorporates a thermal protector however it is not...

Страница 10: ...g Not subject to direct sunlight h No danger of fire i When storing the pump or when operating it within other equipment ensure that the temperature in the vicinity of the pump does not exceed 40 C j...

Страница 11: ...operation Maintenance and Repair 13 Ensure that the pump is discarded in accordance with appropriate legislation related to the treatment and cleaning of waste products Ensure that the pump is discar...

Страница 12: ...ir 9 The pump is manufactured to accurate clearances and should not be dismantled or modified under any circumstances 10 Miscellaneous Problems e g faults with the vacuum pump shall be resolved by con...

Страница 13: ...9 5 DIMENSIONAL DRAWING Fig 1 DA 20D Fig 2 DA 40S...

Страница 14: ...d OFF connect the power cord to the power supply specified the motor rating 7 Turn ON the pump to check that pump is pumping 8 After that turn OFF the power to stop the pump 9 Make piping arrangement...

Страница 15: ...ssure If the pump is used for other applications consult with us 7 2 Startup in Cold Season In a cold season it may sometimes be difficult to start the pump due to hardened bearing grease diaphragm or...

Страница 16: ...nd to the voltage change and frequency change in Domain B it shall operate with main rated values and shall be used practically convenient However operation with it is convenient and safe is maintaine...

Страница 17: ...er s service division 8 2 Maintenance The following checks are required at least once every three days during operation If a problem is found take the measures described in 7 5 TROUBLESHOOTING 1 Check...

Страница 18: ...if hardening crack elongation or the similar is found Air filter Replace the air filter if dirt clogged hardening or the similar is found Bearing If grease runs out or unusual sound or vibrating sound...

Страница 19: ...ad cover Then remove the four hexagon socket head cap screws M6 20 and remove the pump head Remove a cross recessed head machine screws M3 5 remove pump head both sides and replace the suction exhaust...

Страница 20: ...16 Fig 4 Exploded view DA 40S...

Страница 21: ...um pressure Vent pump to atmospheric pressure Height above sea level during storage and installation 1000m or less Unusual sound 1 Bearing damaged Replace 2 Diaphragm broken Replace 3 Motor damaged Re...

Страница 22: ...anual performance specifications maintenance and inspection etc d Malfunctions caused by modifications or repairs carried out by a party other than the manufacturer or by a service company not approve...

Страница 23: ...Industrial Safety and Health Law designates particular substances as the materials to be notified 2 Usage Status Operation Method Approx hours per day years and months Continuous Operation Intermitte...

Страница 24: ...6 6397 2286 ULVAC KIKO Inc https ulvac kiko com en Please contact us for products Service Base or other Inquiries from here ULVAC Inc Components Division 2500 Hagisono Chigasaki Kanagawa 253 8543 Japa...

Отзывы: