Power Supply 

(5) Always check the pump and install wiring with the power cord disconnected. Working with 

power connected may result in electric shock, or the pump starting unexpectedly. 

(6) Ensure that wiring is installed by persons qualified in accordance with regulations for wiring (e.g. 

technical standards for electrical equipment, requirements for internal wiring). Incorrect wiring 

with may result in fire. 

(7) Ensure that lead wiring is properly connected and insulated with insulation tape (or an insulating 

cap). 

(8) This pump has thermal protector. But it is detect of temperature only. Suitable overcurrent 

protection device shall be installed at the final installation.

 

(9) Always ensure that the pump is correctly earthed. A dedicated earth leakage breaker is 

recommended. Failure to earth the pump correctly may result in electric shock if a fault or earth 

leakage occurs. 

(10) Use the pump only at the rated voltage. Use at other than the rated voltage will interfere with 

operation of the overload protection device, and this may result in the motor burning out, or fire. 

(11) Do not damage, modify, and pull the power cord, or place objects on it. Damage to the cord 

may result in electric shock or fire. 

(12) Always fully insert the power cord into the socket. Partial insertion may result in electric shock. 

(13)  Remove the cord from the socket while holding the plug. Failure to do so may result in electric 

shock. 

(14)   Touching the power cord with wet hands may result in electric shock. 

(15)   Touching electrical wiring etc while inserting the power plug may result in electric shock. 

(16) The electric motor fitted to this pump incorporates a thermal protector, however it is not 

guaranteed to be open when it eventually fails at some time after the guaranteed 6000 cycles. 
For safety reasons therefore, it is necessary to fit an overload (over current), protection device 
and an earth leakage breaker. 

Operation 

(17) Inserting fingers or objects into the motor inlet may result in electric shock, injury, or fire. 
(18) This pump is not designed to be explosion-proof. When using the pump, ensure that there are 

no inflammable materials such as solvents, or explosive gases, in the vicinity. Use under such 
conditions may result in injury or fire. 

(19) Do not operate the pump with the outlet blocked, or with equipment which may impede the 

flow of air placed in front of the outlet. Operation under such conditions will result in an 
increase in the internal pressure of the pump, and may result in rupture of the pump itself. 

This pump is not pressure resistant, and may be used at a maximum internal pressure of 30.0 
kPa (gauge pressure). 
Operation at a discharge resistance equivalent to 30.0 kPa (gauge pressure) may result in the 
motor being overloaded. 

(20) Do not use it for the human body.   

And do not use in applications involving organ transplants, or contact with body fluids or living 

tissue. 

(21) Ensure that cooling and load conditions for the particular application are considered 

thoroughly, and that the temperature increase for each motor component is within the 
requirements for class of motor insulation as legally required in your country.

 

 

!

Warning

 

Содержание DA-20D Series

Страница 1: ...100V B 115V C 220V ULVAC KIKO Inc Request to Users Please read this manual thoroughly to ensure safe and effective use of the equipment Keep this manual in a safe place Due to periodic improvements in...

Страница 2: ...9 Compressors and vacuum pumps Safety requirements Part2 Vacuum pumps IEC EN 61010 1 2010 A1 2019 Safety requirement for electrical equipment for measurement control and laboratory use Part1 General r...

Страница 3: ...elated is in conformity with the following standards BS EN 1012 2 1996 A1 2009 Compressors and vacuum pumps Safety requirements Part2 Vacuum pumps BS IEC EN 61010 1 2010 A1 2019 Safety requirement for...

Страница 4: ...ON 10 7 CAUTIONS IN OPERATION 11 7 1 Cautions in Operation 11 7 2 Startup in Cold Season 11 7 3 Thermal Protector 11 7 4 Fluctuations in the power voltage and frequency 12 8 Maintenance Inspection and...

Страница 5: ...quipment correctly Read the section on Safe Use particularly closely Keep This Manual in a Safe Place After reading this manual be sure to keep it in a safe place which is readily accessible to others...

Страница 6: ...outlet caps fitted to inlet and outlet 2 3 Is the pump damaged in any way 4 Are any external screws or inlet and outlet pipes loose Are any components missing Contact your agent or the sales division...

Страница 7: ...uipment may result in death or serious injury to the operator Incorrect handling of the equipment may result in light or medium injuries to the operator or damage to the equipment Incorrect handling o...

Страница 8: ...ir the pump always note the gas which the pump has been used with on the Usage Check Sheet Note that if it has been used to discharge toxic gas for any reason it will be contaminated Please be aware t...

Страница 9: ...ectric shock 15 Touching electrical wiring etc while inserting the power plug may result in electric shock 16 The electric motor fitted to this pump incorporates a thermal protector however it is not...

Страница 10: ...g Not subject to direct sunlight h No danger of fire i When storing the pump or when operating it within other equipment ensure that the temperature in the vicinity of the pump does not exceed 40 C j...

Страница 11: ...operation Maintenance and Repair 13 Ensure that the pump is discarded in accordance with appropriate legislation related to the treatment and cleaning of waste products Ensure that the pump is discar...

Страница 12: ...ir 9 The pump is manufactured to accurate clearances and should not be dismantled or modified under any circumstances 10 Miscellaneous Problems e g faults with the vacuum pump shall be resolved by con...

Страница 13: ...9 5 DIMENSIONAL DRAWING Fig 1 DA 20D Fig 2 DA 40S...

Страница 14: ...d OFF connect the power cord to the power supply specified the motor rating 7 Turn ON the pump to check that pump is pumping 8 After that turn OFF the power to stop the pump 9 Make piping arrangement...

Страница 15: ...ssure If the pump is used for other applications consult with us 7 2 Startup in Cold Season In a cold season it may sometimes be difficult to start the pump due to hardened bearing grease diaphragm or...

Страница 16: ...nd to the voltage change and frequency change in Domain B it shall operate with main rated values and shall be used practically convenient However operation with it is convenient and safe is maintaine...

Страница 17: ...er s service division 8 2 Maintenance The following checks are required at least once every three days during operation If a problem is found take the measures described in 7 5 TROUBLESHOOTING 1 Check...

Страница 18: ...if hardening crack elongation or the similar is found Air filter Replace the air filter if dirt clogged hardening or the similar is found Bearing If grease runs out or unusual sound or vibrating sound...

Страница 19: ...ad cover Then remove the four hexagon socket head cap screws M6 20 and remove the pump head Remove a cross recessed head machine screws M3 5 remove pump head both sides and replace the suction exhaust...

Страница 20: ...16 Fig 4 Exploded view DA 40S...

Страница 21: ...um pressure Vent pump to atmospheric pressure Height above sea level during storage and installation 1000m or less Unusual sound 1 Bearing damaged Replace 2 Diaphragm broken Replace 3 Motor damaged Re...

Страница 22: ...anual performance specifications maintenance and inspection etc d Malfunctions caused by modifications or repairs carried out by a party other than the manufacturer or by a service company not approve...

Страница 23: ...Industrial Safety and Health Law designates particular substances as the materials to be notified 2 Usage Status Operation Method Approx hours per day years and months Continuous Operation Intermitte...

Страница 24: ...6 6397 2286 ULVAC KIKO Inc https ulvac kiko com en Please contact us for products Service Base or other Inquiries from here ULVAC Inc Components Division 2500 Hagisono Chigasaki Kanagawa 253 8543 Japa...

Отзывы: