5.
Open the charge valve slowly and perform the following instruction according to the state of
the compressor.
a.
If the compressor unit is running under normal operating conditions, replenish it with
the pure helium gas until it reaches the operation pressure described in the
compressor instruction manual.
b. If the compressor unit is not running, replenish it with the pure helium gas until it
reaches to the static pressure which is indicated on the front panel.
If helium gas has been charged more than the prescribed pressure of 1.9MPaG or more, the
pressure relief valve on the refrigerator may be going to work. Therefore charge helium gas
slowly so that the pressure relief valve should not operate. On the other hand, the pressure
relief valve in the compressor unit is set at 2.5MPaG.
6. Close the charge valve after charging helium gas.
7. Close the regulator and remove the charging hose from the charge inlet.
The replenishment work of the gas for the compressor has been completed with this.
F gu
High Purity Helium Bottle
Charging Hose
Charging Port
Cooling Water Line
Coldhead Cable
Flexible Hose
Power Cable
Compressor Unit
Remove Front Panel
Cryopump
i re 8-1 Charging Helium Gas
High Pressure
Low Pressure
Bottle Valve
Regulator
SUPPLY
PRESSURE
STS
RUN
CH1
RDY
PWR
COM
RST
8
7
6
5
4
3
2
1
RETURN
SUPPLY
Charge pressure displayed
Содержание CRYO-U Series
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 24: ...This page intentionally left blank S 12 ULVAC CRYOGENICS INCORPORATED...
Страница 26: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 36: ...Cryopump Description 1 10 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 50: ...Installation 4 10 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 64: ...Regeneration 6 8 ULVAC CRYOGENICS INCORPORATED This page intentioanlly left blank...
Страница 68: ...Disconnection and Storage 7 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 82: ...Maintenance 8 14 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 90: ...Troubleshooting A 8 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 92: ...Flexible Hose B 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 104: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...