①
Hold the braid support straight and tighten the self-sealing coupling
by hand.
②
Tighten the self-sealing coupling using two wrenches
until the
fittings are firmly sealed.
When using torque wrench, the
recommended torque value is 20N
・
m.
Figure 4-5 Connecting Flexible Hoses
Braid Support
Tighten with a
wrench.
Hold with a wrench.