SonixMDP/SP/OP Service Manual 00.053.101, Revision C
Chapter 7: Field Service Components
7-19
5.
Using the 7mm wrench, undo the nut and washer securing the ground wire from the modulo to the LCD display
and, if applicable, to the ECG (refer to image in
).
6.
Disconnect the three (3) (bottom left rear) power cables connected to the modulo: LCD display power cable
(
1
), UPS input power cable (
2
) and UPS output power cable (
3
).
7.
Using the 4mm Allen key, undo the two (2) screws and washers securing the modulo to the front of the system
frame.
8.
Using the #2 Phillips screwdriver, undo the four (4) screws securing the modulo to the side of the system
frame.
9.
Pull the modulo out of the system frame, then lift it clear.
Note:
The LCD display power cable (
1
) is color coded with
green
tape.
System Power Connection
EMI Filter
1
2
3
1
2
3
M4x8 Pan Head Screw (x4)
Tool: #2 Phillips screwdriver
M5 Split Lock Washer (x2)
M5x12 Socket Head Screw (x2)
Tool: 4mm Allen key
Содержание SonixMDP
Страница 1: ...SonixMDP SP OP Ultrasound System Service Manual...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of Contents 00 053 101 Revision C SonixMDP SP OP Service Manual vi...
Страница 12: ...List of Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual ii...
Страница 44: ...Chapter 4 Performance Testing 00 053 101 Revision C SonixMDP SP OP Service Manual 4 4...
Страница 48: ...Chapter 5 Software 00 053 101 Revision C SonixMDP SP OP Service Manual 5 4...
Страница 124: ...Chapter 9 Network Configuration 00 053 101 Revision C SonixMDP SP OP Service Manual 9 8...
Страница 142: ...Chapter 11 Maintenance 00 053 101 Revision C SonixMDP SP OP Service Manual 11 16...
Страница 150: ...Chapter 12 Troubleshooting Issues 00 053 101 Revision C SonixMDP SP OP Service Manual 12 8...
Страница 166: ...Appendix C Ultrasonix Limited Warranty 00 053 101 Revision C SonixMDP SP OP Service Manual C 2...
Страница 168: ...Appendix D Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual D 2...