ROTATING/LOCKING THE CLEAVER BLADE
Blade rotating mode.
Initial setting mode will be set as shown picture (Rotation mode).
Adjust gear pusher position after loosening Setscrew as shown picture.
Blade locking mode.
When the blade is used without rotating.
ELECTRODE REPLACEMENT
The K33A electrodes are designed to last for 18,000 arc cycles. Once 18,000 arc cycles are
reached the a “Replace Electrodes” message will be displayed. Once the number of arc
discharges exceeds the threshold the results could lead to higher splice loss and weaker
tensile force at the splicing point. Each new K33A comes with a spare pair of the EI-23
electrodes included in the standard kit.
Turn the power off when replacing electrode.
Remove the electrode block and the electrode.
Clean the electrode carefully using soft cotton swab moistened with alcohol
and then install them into the electrode blocks per illustration below.
Replace the electrode blocks into the fusion K33A.
Turn the power on and conduct electrode stabilizing.